CONFIGURATION AND ANALYSIS SOFTWARE ENSUITE
Information for general use
Rev. M / 73023639
220
GasLab Q2
8.6.7.
Use “Live data tab” to view Time
Service data
Information around time setting, related events and data can be obtained, if the time service export
values are used in the tab
“Live data
” as shown below.
Value
Meaning
Current time
System time
Latest update
Time stamp of the last time change.
Latest deviation
Deviation (in seconds) at the last time change
Latest external time source
Number of the external source for the last time synchronization
Daylight saving time
(DST) is the practice of advancing clocks during warmer months
Figure 8.54: Time service main data
.
The
Time zone
(geographical location)
and the setting of DST could be viewed in the parameter settings of Basic system.
See next figure.