GASLAB Q2 DISPLAYS (OPERATE THE DEVICE)
Information for general use
Rev. M / 73023639
126
GasLab Q2
7.3.15.
Display Q2 sensor values
The list of sensor values can be opened by clicking on this symbol.
Generally, this display is only used for service purposes and for locating errors.
Figure 7.41:
Q2 sensor values display
A complete list of all sensor values is displayed in the following table:
Sensor signal
Use/Meaning
Sensor value range
UIR 1
For internal calculations only
Several 100 mV
UIR 2
For internal calculations only
UIR 3
For internal calculations only
UIR 4
For internal calculations only
PI
Pressure (a measuring unit [approx.
160 mbar])
Ambient-dependent
TSB
Temperature of sensor block
approx. 70°C
≙
158°F
≙
343K
UW11
For internal calculations only
Several 100 mV
UW12
For internal calculations only
UW21
For internal calculations only
UW22
For internal calculations only
PU
Ambient pressure (internal monitoring)
Ambient-dependent
TP
Circuit board pressure (internal monitoring)
Ambient-dependent
PW1
For internal calculations only
Several mW
PW2
For internal calculations only
TM1
For internal calculations only
Around 100°C
TM2
For internal calculations only
TB1
For internal calculations only
Around 68°C
TB2
For internal calculations only
TH1
Housing temperature (internal monitoring)
Ambient-dependent