GASLAB Q2 DISPLAYS (OPERATE THE DEVICE)
Information for general use
Rev. M / 73023639
104
GasLab Q2
Step
Change a TLS certificate and publish it to enSuite
1.
Log in to the "
User Main Display
" on the device or on the remote operationl panel
(Home >> System >> Users).
See also section
7.3.11 System display Users (Login/Logout / Password)
2.
Switch to Diplay
Info (System)
via
Home
>. Activate the <device serial number >.
You will see the 2nd page of the display as shown below.
3.
Delete the existing certificate with the action
Delete certificate
.
The text of the action will change, see figure.
The certificate will only be permanently deleted on shutdown
4.
Restart the device with the action
Reboot and create certificate
. On restart, the
device generates a new certificate for MMS communicationt.
7.3.4.
Info display (Device monitor)
The device monitor shows information about
the operating hours and battery charge and
about the CPU and RAM load.
Figure 7.19:
System info
–
Device monitor display
The battery is mainly used when the device is switched off. After service or repair work (including
battery replacement), a user with the appropriate rights can reset the battery status to 100%.
The battery must be replaced when the charge falls to below 20%. A discharged battery
may, in certain circumstances, result in the loss of calculated and saved measurement
data. The device will no longer start correctly.
DANGER OF EXPLOSION IF THE BATTERY IS NOT REPLACED CORRECTLY.
The battery replacement must be carried out by Honeywell service personnel.