GasAlertMicro
User Options Menu
If a sensor option is enabled on the LCD but the new sensor
is not installed on the detector, the following message
displays:
H2S SEnSor not inStALLEd - SEnSor CAnnot bE
EnAbLEd
.
3. Use
G
or
H
to scroll to a new user option, or press
C
to exit user options and return to normal
operating mode.
Pass Code Protection
The
PASS Lock
option is used to prevent access to the user
options and calibration/alarm setpoint procedure without a
pass code.
Note
For calibration, the pass code is requested after
auto zero is performed (if enabled).
If the detector is pass code protected, the key icon (
)
displays at the bottom of the LCD during start-up.
Note
The detector is shipped with the pass code
protection option disabled.
Enable/Disable Pass Code Protection
To enable/disable the pass code protect option, complete the
following:
1. From the user options menu, use
G
or
H
to scroll
to the
PASS Lock
option.
2. Press
C
to select this option.
The detector emits two beeps and the LCD displays
if pass code protection is enabled or disabled.
or
The LCD displays the current mode (
On
or
OFF
)
and then returns to the
FiniSh OPtionS
screen.
Note
The pass code is provided on a separate card
inside the shipping container.
25