AINTENANCE
8.8
ThinPrep™ 5000 Processor with AutoLoader Operator’s Manual
8
Figure 8-7 Waste Cabinet Door Open
On a weekly basis pull out the filter waste bin and empty it of used filters. Depending on the
processing volume of your laboratory, the filter waste bin may require more or less frequent
emptying. Used filters may be disposed of as regular waste.
If desired, the waste bin may be wiped down with soap and water, or a 10% bleach solution but be
sure to allow it to dry thoroughly before re-use.
Return the waste bin to the location it was originally in, making sure it is positioned to receive filters
from the waste chute directly above it.
The slide waste bin is in the same compartment, to the left of the filter waste bin. (See Figure 8-8.) The
liquid waste bottle may be in front of it, and it can be moved out of the way to access the slide waste
bin. Remove the waste bin and dispose of slides in an appropriately labeled and enclosed Sharps
container.
If desired, the slide waste bin may be wiped down with soap and water, or a 10% bleach solution but
be sure to allow it to dry thoroughly before re-use.
Return the waste bin to its location and make sure the liquid waste bottle has been returned, as well.
Filter waste chute
Filter waste bin
Slide waste bin is
located behind the liquid
waste bottle
Содержание ThinPrep 5000
Страница 1: ...Operator s Manual ThinPrep 5000 Processor with AutoLoader...
Страница 5: ...This page intentionally left blank...
Страница 6: ...Instructions For Use Instructions For Use...
Страница 43: ...Table of Contents Table of Contents...
Страница 48: ...1 Introduction 1 Introduction...
Страница 73: ...2 Installation 2 Installation...
Страница 82: ...3 PreservCyt and CytoLyt Solutions 3 PreservCyt and CytoLyt Solutions...
Страница 98: ...4 Gynecologic Sample Preparation 4 Gynecologic Sample Preparation...
Страница 107: ...5 Non Gynecologic Sample Preparation 5 Non Gynecologic Sample Preparation...
Страница 134: ...6 User Interface 6 User Interface...
Страница 216: ...USER INTERFACE 6 82 ThinPrep 5000 Processor with AutoLoader Operator s Manual 6 This page intentionally left blank...
Страница 217: ...7 Operating Instructions 7 Operating Instructions...
Страница 256: ...8 Maintenance 8 Maintenance...
Страница 265: ...ThinPrep 5000 Processor with AutoLoader Operator s Manual 8 9 8 MAINTENANCE Figure 8 8 Slide Waste Bin...
Страница 282: ...MAINTENANCE 8 26 ThinPrep 5000 Processor with AutoLoader Operator s Manual 8 This page intentionally left blank...
Страница 283: ...9 Troubleshooting 9 Troubleshooting...
Страница 308: ...10 Staining and Coverslipping 10 Staining and Coverslipping...
Страница 313: ...11 ThinPrep Pap Test Training Program 11 ThinPrep Pap Test Training Program...
Страница 316: ...Service Information Service Information...
Страница 319: ...Ordering Information Ordering Information...
Страница 326: ...Index Index...
Страница 345: ...Operator s Manual ThinPrep 5000 Processor with AutoLoader...