5.4
ThinPrep™ 5000 Processor with AutoLoader Operator’s Manual
5
than one centrifuge tube of sample is used, the cell pellets can be combined after pouring off the
supernatant.
Other Sample Types
For non-mucoid brushings and scrapings that are received in PreservCyt
™
Solution, the sample is
ready to be run on the ThinPrep
™
5000 processor with AutoLoader.
For non-mucoid brushing and scrapings that are received in CytoLyt Solution, follow the protocol for
FNA samples. See page 5.14.
Urine Sample for use with the Vysis
®
UroVysion Assay
Do not exceed a 2:1 ratio of urine to PreservCyt Solution. If the urine volume exceeds 60 mL,
pour off excess. A minimum volume of 33 mL of urine is required to perform the Vysis
®
UroVysion assay.
Other Collection Media
In cases where CytoLyt Solution is contraindicated, balanced electrolyte solutions, such as
Plasma-Lyte and Polysol, may be used as collection media for samples to be processed on the
ThinPrep 5000 processor with AutoLoader. These solutions are primarily used as media for
washings or lavages which contact the patient.
Non-Recommended Collection Media
Hologic does not recommend the use of the following collection solutions with the ThinPrep system.
Use of these solutions will produce sub-optimal results:
•
Sacomanno and other solutions containing carbowax
•
Alcohol
•
Mucollexx
®
•
Normal Saline
•
Culture media, RPMI Solution
•
PBS
•
Solutions containing formalin
Specimens
must
be centrifuged and washed in CytoLyt
™
Solution and transferred to PreservCyt
™
Solution prior to being processed on the ThinPrep
™
5000 processor with AutoLoader.
Содержание ThinPrep 5000
Страница 1: ...Operator s Manual ThinPrep 5000 Processor with AutoLoader...
Страница 5: ...This page intentionally left blank...
Страница 6: ...Instructions For Use Instructions For Use...
Страница 43: ...Table of Contents Table of Contents...
Страница 48: ...1 Introduction 1 Introduction...
Страница 73: ...2 Installation 2 Installation...
Страница 82: ...3 PreservCyt and CytoLyt Solutions 3 PreservCyt and CytoLyt Solutions...
Страница 98: ...4 Gynecologic Sample Preparation 4 Gynecologic Sample Preparation...
Страница 107: ...5 Non Gynecologic Sample Preparation 5 Non Gynecologic Sample Preparation...
Страница 134: ...6 User Interface 6 User Interface...
Страница 216: ...USER INTERFACE 6 82 ThinPrep 5000 Processor with AutoLoader Operator s Manual 6 This page intentionally left blank...
Страница 217: ...7 Operating Instructions 7 Operating Instructions...
Страница 256: ...8 Maintenance 8 Maintenance...
Страница 265: ...ThinPrep 5000 Processor with AutoLoader Operator s Manual 8 9 8 MAINTENANCE Figure 8 8 Slide Waste Bin...
Страница 282: ...MAINTENANCE 8 26 ThinPrep 5000 Processor with AutoLoader Operator s Manual 8 This page intentionally left blank...
Страница 283: ...9 Troubleshooting 9 Troubleshooting...
Страница 308: ...10 Staining and Coverslipping 10 Staining and Coverslipping...
Страница 313: ...11 ThinPrep Pap Test Training Program 11 ThinPrep Pap Test Training Program...
Страница 316: ...Service Information Service Information...
Страница 319: ...Ordering Information Ordering Information...
Страница 326: ...Index Index...
Страница 345: ...Operator s Manual ThinPrep 5000 Processor with AutoLoader...