SMI-1414-EN Rev. 005 - Installation and Service Manual
Page 29 of 51
Figure 23: Confirmation of 16ch Sentinelle coil protocols installed
5. If the Sentinelle table configurations and protocols are on the system skip to the finalization steps in
section 7.4.5.
6. To install the Sentinelle table configurations and protocols, insert the “16Ch Breast Coils by
Sentinelle Coil Definitions Disc” into the default (typically upper) host DVD drive.
7. Open a C-Shell from the Service Desktop manager and type the following commands:
su <enter>
<root password>
mount /mnt/cdrom <enter>
/mnt/cdrom/autorun <enter>
8. A Coil Data Disk Install window will open. Type “i” to install the CD contents.
9. When the content of the CD installation is complete, the cursor will appear and you will see the
message as shown in Figure 26. Press enter to exit the Coil CD installation window.
Note: The number(s) following
the “24.” will differ by system.