Cenova Image Analytics Server 3.0 User Guide
Chapter 2: System Description
Page 10
MAN-05463-001 Revision 002
•
Image objects sent to Cenova by Hologic 3D Mammography
TM
systems known as
Raw Synthesized 2D
images.
•
Image objects sent to Cenova by any of the Hologic film-digitizing systems known as
For Presentation
images.
Some DICOM image objects are not processed by the software, depending upon the
application and mammography view shown in the image.
For more information on image formats, see the diagrams in
on page 12.
2.1.3 Input Ports and Application Licenses
Each device that produces digital mammography image files maps to an input port on
the Cenova server. The DICOM header for each image object includes an identifier (e.g., a
serial number) that indicates the device from which the image originated.
For each Cenova server, sites can support:
•
Up to four licenses for any or all of the 2D software applications plus a single license
for DigitalNow HD, or
•
A single license for DigitalNow HD
Note that exceeding these guidelines with a single server can affect system response
times.
Each Cenova software application requires one software license for each input device.
For example, if there are four FFDM devices, then the Cenova server must be configured
with four software licenses for each desired Cenova software application.
The Cenova server can also process images sent from any number of PACS devices, as
long as the images originated from one of the licensed input devices.
For assistance with license configurations, contact your Hologic Account Manager.