System Operation
27
en
representing the actual deficit volume until the deficit limit is reached (see sec-
tion
Deficit Limit
in Chapter 7, Safety Functions).
5.12
Using the Pump during Surgery
Open bag clamps on the fluid bags (
Fully open hysteroscope inflow stopcock.
If drainage stopcock is available: Fully close drainage stopcock.
Keep the hysteroscope at the height of the patient and above the drape to
ensure the fluid can be collected. Do not insert the hysteroscope into the
uterus at this time.
Press the button
Prime
(
(17)
Fig. 5-1).
Pump will run for approximately 20 seconds to purge air from tubes and run
the automatic lumen calibration function.
Pump will display
Calibration Running.
Automatic lumen calibration
The pump is equipped with an automatic lumen calibration function. The system
determines the flow resistance of the hysteroscope. This resistance is used to cal-
culate the pump pressure necessary to maintain the pre-set intrauterine pres-
sure. To overcome this resistance, the pump allows a pressure of up to 80 mmHg
during calibration. This will be shown in the display of the actual intrauterine
pressure. If calibration fails due to high resistance, calibration is repeated with a
permissible pressure of up to 150 mmHg. If calibration is then still not completed,
the pump displays
Prime Fail - Open Stopcock Clamps
.
The automatic lumen calibration starts once the
Prime
button is pressed.
Three beeps are heard once the automatic lumen calibration is finished. The
pump display will show
Prime Successful Close Stopcock
for 5 seconds fol-
lowed by
System Operating.
Close hysteroscope inflow stopcock to stop fluid flow. Once all fluid has been
removed from the drape, zero the deficit display.
Check to see if fluid has leaked in the area of the pressure chamber. If you
find leaked irrigation fluid in the area of the pump, change the tube set and
retry the automatic lumen calibration.
NOTE!
The calibration must be performed outside of the patient to ensure a correct lu-
men calibration and deficit calculation.
NOTE!
The pump continues to operate after automatic lumen calibration is complete.
The pump should be stopped by closing hysteroscope inflow stopcock.
CAUTION!
A new lumen calibration must be carried out with each change of the hystero-
scope during surgery (5.15, Changing Instrument during Surgery).
System Operation
Open stopcock and guide the hysteroscope with fluid flowing into the uter-
us.
Adjust intrauterine pressure setting as necessary to obtain adequate disten-
sion and visualization.
Completing System Operation
After the medical procedure, if system operation is complete close the hys-
teroscope inflow stopcock.
Wait until the entire fluid volume from the drape and the tube set has been
collected in both containers.
Press the
Pause/Resume
button.
Note the deficit volume indicated on the deficit display. This is the total fluid
volume that was absorbed by the patient.
Содержание Aquilex
Страница 4: ......
Страница 8: ......
Страница 10: ......
Страница 61: ...Appendix 53 en 16 Appendix 16 1 Test Log Date Tests Performed Results Comment Signature ...
Страница 64: ......
Страница 118: ...56 Anhang de 16 Anhang 16 1 Testprotokoll Datum Durchgeführte Tests Ergebnis Bemerkung Unterschrift ...
Страница 120: ......
Страница 122: ......
Страница 176: ...56 Annexe fr 16 Annexe 16 1 Procès verbal de tests Date Tests réalisés Résultat Remarque Signature ...
Страница 178: ......
Страница 180: ......
Страница 234: ...56 Apéndice es 16 Apéndice 16 1 Registro de test Fecha Test realizados Resultado Anotaciones Firma ...
Страница 236: ......
Страница 238: ......
Страница 292: ...56 Appendice it 16 Appendice 16 1 Protocollo di verifica Data Test eseguiti Risultato Nota Firma ...
Страница 294: ......
Страница 296: ......
Страница 351: ...Bijlage 57 nl 16 Bijlage 16 1 Testprotocol Datum Uitgevoerde tests Resultaat Opmerking Handtekening ...
Страница 353: ......
Страница 354: ......