Affirm User Guide
Chapter 4:
How to Use the System
MAN-03209 Revision 005
Page 39
Biopsy Staging
The buttons on the Biopsy Staging area communicate target information to the Biopsy
Control Module. The fields on the right side of the buttons show the selected target and
biopsy device.
Figure 31:
Function Buttons and Data on the Biopsy Tab
Figure Legend
1.
Create Target
sends the current target list that shows in the Biopsy Staging area to the
Biopsy Control Module.
Caution:
If the Biopsy Control Module is unplugged after the
targets are transmitted, the Biopsy Control Module deletes
the targets. Resend the targets.
2.
Reject Target
removes the selected target from the list that displays in the Biopsy
Staging area, if that target was not created.
3.
Resend Target
resends the selected target set to the Biopsy Control Module.
4.
Project Target
shows the selected target on an additional stereo pair on the Preview
screen.
5.
Delete Target
deletes the selected target if that target was created.
6.
Move Z-Target Up
moves the final position of the needle away from the Breast
Platform and the graphic of the lesion downward. The values for the safety margins
change accordingly.
Содержание Affirm
Страница 1: ......
Страница 2: ......
Страница 3: ...User Guide For Software Version 1 7 Part Number MAN 03209 Revision 005 January 2015...
Страница 4: ......
Страница 8: ......
Страница 10: ......
Страница 16: ......
Страница 24: ......
Страница 32: ......
Страница 54: ......
Страница 58: ......
Страница 62: ......
Страница 64: ......
Страница 66: ......
Страница 69: ......
Страница 70: ......