Affirm User Guide
Chapter 3:
How to Install or Remove the System
MAN-03209 Revision 005
Page 19
3.2.3 Needle Guides
Warning:
Always use sterile techniques when you use Needle
Guides during the patient procedures.
Warning:
It is important to install the device correctly. Be sure
to insert the needle through the top and bottom
Needle Guides.
To install a disposable Needle Guide:
1.
Align the Needle Guide so that the raised-square side of the Needle Guide fits
between the two lobes of the Needle Guide Mount.
2.
Slide the open area of the U-shape in the Needle Guide around the pin in the Needle
Guide Mount.
3.
Push the Needle Guide in until it locks into position.
Figure 9:
How to Install the Needle Guides
Figure Legend
1. Needle Guide
2. Needle Guide Mount
Note
The Needle Guides can look different from the Needle
Guide shown.
To remove a disposable Needle Guide:
1.
Remove the biopsy device from the Z-axis Slide Rail.
2.
Pull the Needle Guide away from the pin and remove from the Needle Guide Mount.
3.
Discard the Needle Guide in accordance with local regulations.