Affirm Instructions for Use
Chapter 3
—
How to Use the System
System Verifications
Part Number MAN-02866
15
Chapter 3
How to Use the System
1.0 System
Verifications
1.1
Confirm the Host Connection
When
the
Selenia
Dimensions
is
On
and
the
Affirm
cable
connections
are
correct,
the
Home
screen
displays
on
the
Biopsy
Control
Module.
Figure
10:
Home
Screen
on
the
Biopsy
Control
Module
1.2
Perform the QAS Needle Test
Perform
this
test
one
time
each
day
you
plan
to
use
the
system
to
confirm
the
system
accuracy.
Record
your
results
in
the
QAS
Needle
Test
Checklist
on
page
Caution:
Do
not
extend
the
QAS
Needle
unless
the
needle
is
attached
to
the
Biopsy
Guidance
Module,
and
the
module
is
installed
on
the
C
‐
arm.
You
can
use
Auto
Biopsy
Mode
or
Manual
Biopsy
Mode
for
the
QAS
Needle
Test.
The
two
sections
that
follow
describe
each
method.
For
more
information
about
biopsy
modes,
see
Stereo
Biopsy
Modes
on
page
28.
Содержание Affirm MAN-02866
Страница 1: ......
Страница 2: ......
Страница 3: ...Instructions for Use For Software Version 1 6 Part Number MAN 02866 Revision 002 May 2012...
Страница 4: ......
Страница 8: ......
Страница 10: ......
Страница 16: ......
Страница 24: ......
Страница 52: ......
Страница 56: ......
Страница 60: ......
Страница 62: ......