HP-90E 

DIGITAL MULTIMETER 

OPERATOR’S MANUAL 

1.

Overview

The multimeter is characterized at slim size, portable, stable performance 

and  anti-dropping  capacity.  Using  4000  counts  digit  LCD  monitor  with 

character  22mm  high,  they  offer  clear  readings.    With  overall  circuitry 

design  centering  on  large-scale  IC  A/D  converters  in  conjunction  and 

over-load  protection  circuit,  the  meters  give  excellent  performance  and 

exquisite making as a handy utility instrument. 

The meters can be used to measure DC & AC voltage, DC & AC current, 

resistance, capacitor, frequency, duty cycle, temperature, battery test, 

positive diode voltage fall and audible continuity. 

2. Panel Layout

LCD display: 4000 counts digit, full function symbol display.

SELECT key: This key work on the "

" range, press the key

to choose resistance, diode or continuity test, on the voltage or

current range, change to DC or AC; If press and hold SELECT key

to power on, “Auto Power Off ” function will be disabled.

HOLD

 key: Press the “

HOLD

” key to lock display value, and the “

DH

” 

sign will appear on the display, press it again to exit.

RANGE

  Key:  Press  the  “

RANGE

”  key,  the  meter  enters  manual

range  mode,  press  it  more  than  2  seconds  again,  return  to  auto

mode.

REL

 Key: Press the “

REL

” key, the meter enters relative measuring

mode,  “

REL

”  is  displayed  on  the  LCD  and  the  present  reading

becomes the reference value and displayed on the display. Relative

measurement R

EL

=measurement value-Reference value

.

Hz/%

 Key: On “ACV/ACA” or “Hz” range, press the “

Hz/%

” key, you

can choose the 

Frequency

 or 

Duty Cycle

 measurement.

CDS  sensor:  The  CDS  sensor  can  reaction  to  the  ambient

brightness range, then automatically control the LCD backlight

to lighten or go out.

 

Rotary Switch: use this switch to select functions and ranges.

VHz

  Input Jack 

COM and Temperature “-” Input Jack 

 

μAmA/BATT and Temperature “+” Input Jack 

 

20A Input Jack 

3. Safety Information

3-1 The meters are designed according to IEC-1010 concerning electronic 

measuring instruments with an over-voltage category 600V (CAT 

) and 

pollution 2. 

3-2 Follow all safety and operating instructions to ensure that the meter is 

used safely and is kept in good operating condition. 

3-3 safety symbols: 

Important safety information, refer to the operating manual. 

Dangerous voltage may be presence. 
Double insulation (protection Class II)

 

4. Special Cautions for Operation

4-1 The meters can be safe only according to standard procedures when 

used in conjunctions with the supplied test leads. To replace damaged test 

leads with only the same model or same electric specifications. 

4-2 To avid risk of electric shock, do not use the meters before the cover is 

in place. 

4-3 The range switch should be right position for the testing. 

4-4  To  avoid  electric  shock  and  damaging  the  instruments,  the  input 

signals are forbidden to exceed the specified limits. 

4-5 When measuring TV set or switched power, attention should be paid to 

the possible pulses that may bring destruction to the circuit. 

4-6  Range  switch  position  is  forbidden  to  be  changed  at  random  during 

measurement. 

4-7 Take caution against shock in the course of measuring voltage higher 

than DC 60V & AC 30V. 

4-8  Protection  fuse  should  be  replaced  only  with  same  type  and  same 

specification. 

4-9 After operation is finished, set function switch  at OFF range to save 

battery power. 

4-10 If the meter is without usage for long time, take out battery to avoid 

damage by battery leakage. 

5. GENERAL SPECIFICATIONS

5-1 Max Voltage between input terminal and Earth Ground: CAT 

 600V 

5-2 Over-range Indication: display “OL” for the significant digit. 

5-3 Automatic display of negative polarity “-” . 

5-4 Low Battery Indication: “ ” displayed. 

5-5 Max LCD display: 4000 counts digit. 

5-6 Auto range & Manual range control 

5-7  Auto  Power  Off:  When  measurement  exceeds  30  minutes  without 

switching  mode  and  pressing  key,  the  meter  will  switch  to  standby 

mode. Press any key to exit standby mode. When restart the system, 

press and hold 

SELECT

 key to disable auto power off. 

5-8 Auto LCD backlight 

5-9 Fuse protection: 400mA/250V PPTC Resettable Fuse 

F-20A/250V Fuse (Ø5x20mm) 

5-10 Power supply: 9V battery (6F22) 

5-11 Operating Temp.: 0

  to 40

  (relative humidity <85%) 

5-12 Storage Temp.:-10

  to 50

  ((relative humidity <85%) 

5-13 Guaranteed precision Temp.: 23

±

5

  (relative humidity <70%) 

5-14 Dimension: 193x88x41mm 

5-15 Weight: approx. 320g (including battery)

 

6. Testing Specifications

Accuracy is specified for a period of year after calibration and at 18

  to 

28

  (64

  to 82

) with relative humidity to 70%. 

6-1 DC Voltage

 

Range 

Resolution 

Accuracy 

400mV 

0.1mV 

±(0.5% of rdg + 2 digits) 

4V 

1mV 

40V 

10mV 

400V 

100mV 

1000V 

1V 

±(0.8% of rdg + 2 digits) 

-- Impedance: 10MΩ, More than 100MΩ on 400mV range

 

--  Overload  protection:  250V  DC  or  AC  rms  for  400mV  range,  effective 

1000V DC or 750V AC rms for other ranges 

6-2 AC Voltage 

Range 

Resolution 

Accuracy 

400mV 

0.1mV 

±(1.2% of rdg + 3 digits) 

4V 

1mV 

±(0.8% of rdg + 3 digits) 

40V 

10mV 

400V 

100mV 

750V 

1V 

±(1.2% of rdg + 3 digits) 

--The 400mV range be selected by press “

RANGE

” key only 

-- Impedance: 10MΩ, More than 100MΩ on 400mV range

 

--  Overload  protection:  250V  DC  or  AC  rms  for  400mV  range,  effective 

1000V DC or 750V AC rms for other ranges 

-- Frequency Range: 40 to 400Hz 

-- Response: average, calibrated in rms of sine wave 

6-3 DC Current 

Range 

Resolution 

Accuracy 

4

00μA 

0.1

μA 

±(1.2% of rdg + 2 digits) 

4

000μA 

1

μA 

40mA 

10

μA 

400mA 

100

μA 

4A 

1mA 

±(2.0% of rdg + 3 digits) 

20A 

10mA 

-- Overload protection: 400mA/250V PPTC Resettable Fuse 

F-20A/250V Fuse, 20A up to 10 seconds 

6-4 AC Current 

Range 

Resolution 

Accuracy 

4

00μA 

0.1

μA 

±(1.5% of rdg + 3 digits) 

4

000μA 

1

μA 

40mA 

10

μA 

400mA 

100

μA 

4A 

1mA 

±(2.5% of rdg + 5 digits) 

20A 

10mA 

-- Overload protection: 400mA/250V PPTC Resettable Fuse 

F-20A/250V Fuse, 20A up to 10 seconds 

-- Frequency Range: 40 to 400Hz 

-- Response: average, calibrated in rms of sine wave 

6-5 Resistance 

Range 

Resolution 

Accuracy 

4

00Ω 

0.1Ω 

±(1.0% of rdg + 3 digits) 

4

kΩ 

1Ω 

±(1.0% of rdg + 2 digits) 

4

0kΩ 

10Ω 

4

00kΩ 

100Ω 

4

MΩ 

1kΩ 

4

0MΩ 

10k

Ω 

±(1.5% of rdg + 3 digits) 

-- Overload protection: 250V DC or AC rms 