3 1/2 DIGITAL DOUBLE OPEN JAW 

AC CLAMP MULTIMETER

 

OPERATION MANUAL 

1. SAFETY INFORMATION

 

The following safety information must be observed to insure 
maximum personal safety during the operation at this meter. 
1.1 Do  not  operate  the  meter if the body of    meter or the 
test lead look broken. 
1.2 Check    the main function    dial and make sure it is at the 
correct position before each    measurement. 
1.3 Do  not  perform  resistance  test on a live power system. 
1.4  Do  not  apply  voltage  between  the  test terminals  
and test terminal to ground that exceed the maximum    limit   
record in this manual. 
1.5 Exercise  extreme  caution  when measuring live  system  
with voltage    greater than 60V DC or 30V AC. 
1.6 Change  the battery  when  the   " "      symbol   
appears to avoid incorrect data. 

2. Panel Layout 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

(1) Clamp jaws

Opens 38mm to enclose conductor. 

(2), (3) Jaw-opening handle

Opens and closes the jaws. 

(4) Rotary Switch

:

 use this switch to select functions and ranges.

 

(5) D.HOLD key

In any range, push the key, the present display 

value will be locked and the "DH" symbol will appear, push it 
again to exit HOLD and the "DH" symbol disappear. 

(6) CDS sensor: The CDS sensor can reaction to the ambient 
brightness range, then automatically control the LCD 
backlight to lighten or go out.

 

(7) LCD Display

 

(8) 

 Input Jack. 

(9) COM Input Jack. 
(10) V Input Jack.

 

3. SPECIFICATIONS 
3.1 GENERAL SPECIFICATIONS 

Display:    3 1/2 digit LCD with a max. reading of 1999. 
Polarity: Automatic negative polarity indication. 
Zero adjustment: Automatic. 
Over-range indication: Only the "1" or "-1" display. 
Low battery indication:  " " 
Safety Standards: The  meter  is up  to  the  standards  of 
IEC1010 Double    Insulation, Pollution    Degree 2, Over-voltage 
Category 

Data hold: Push the D.HOLD key and the DH sign will appear on 
the display. 
Clamp opening size: 38mm. 
Operating Environment: 
Temperature 32

104

(0

40

), humidity

80%RH. 

Storage Environment: 
Temperature -4

140

(-20

60

), humidity

90%RH. 

Power supply: 3×1.5V AAA batteries. 
Dimension: 193(H)×73(W)×26(D)mm. 
Weight: Approx. 325g (including battery). 

3.2 ELECTRICAL SPECIFICATIONS 

Accuracies are ± (% of r number in last digit) 
at 23

±

5

,

75%RH. 

3.2.1 DC Voltage 

Range Accuracy 

Resolution 

500V 

0.8% of rdg + 2 digits 

1V 

Overload protection: 500V DC / 500V rms AC 
Impedance: 10M

 

3.2.2 AC Voltage 

Range Accuracy 

Resolution 

500V 

1.2% of rdg + 3 digits 

1V 

Average sensing, calibrated to rms of sine wave 
Frequency: 40

400Hz 

Overload protection: 500V DC / 500V rms AC 
Impedance: 10M

 

3.2.3 AC Current 

Range Accuracy 

Resolution 

 

20A 10mA 

200A 

2.0% of rdg + 5 digits 

0.1A 

400A 

2.5% of rdg + 5 digits 

1A 

Average sensing, calibrated to rms of sine wave 
Frequency: 50

60Hz 

Overload protection: AC 400A within 60 seconds 