HP-37A

DIGITAL MULTIMETER

OPERATOR’S MANUAL

1.

Overview

The multimeter is characterized at slim size, portable, stable performance

and anti-dropping capacity. Using 3 1/2 digits LCD monitor with character

28mm high, they offer clear readings.

With overall circuitry design

centering on large-scale IC A/D converters in conjunction and over-load

protection circuit, the meters give excellent performance and exquisite

making as a handy utility instrument.

The meters can be used to measure DC & AC voltage, DC current,

resistance, temperature, battery test, positive diode voltage fall and

audible continuity.

2. Panel Layout

1

Test lead fixture: Fix the test lead.

2

LCD display: 3 1/2 digits LCD display.

3

HOLD

key: Press the “

HOLD

” key to lock display value, and the “

DH

sign will appear on the display, press it again to exit.

4

LIGHT

key: Press the “

LIGHT

” key to light up the LCD backlight,

press it again to exit, it will be automatically turned off too after

approx. 30 seconds. If the battery is in weak power, the light will be

dimmed. Press the key to power on when auto power off.

5

Rotary Switch: Use this switch to select functions and ranges.

6

COM:

COM and Temperature “-” Input Jack

7

10A:

10AInput Jack

8

V

mA:

V/mA/BATT/

and Temperature “+” Input Jack

9

Crust of meter

10

Protective casing

3. Safety Information

3-1 The meter is designed according to IEC-1010 concerning electronic

measuring instruments with an over-voltage category 600V (CAT

) and

pollution 2.

3-2 Follow all safety and operating instructions to ensure that the meter is

used safely and is kept in good operating condition.

3-3 safety symbols:

Important safety information, refer to the operating manual.

Dangerous voltage may be presence.
Double insulation (protection Class II)

4. Special Cautions for Operation

4-1 The meters can be safe only according to standard procedures when

used in conjunctions with the supplied test leads. To replace damaged test

leads with only the same model or same electric specifications.

4-2 To avid risk of electric shock, do not use the meters before the cover is

in place.

4-3 The range switch should be right position for the testing.

4-4 To avoid electric shock and damaging the instruments, the input

signals are forbidden to exceed the specified limits.

4-5 When measuring TV set or switched power, attention should be paid to

the possible pulses that may bring destruction to the circuit.

4-6 Range switch position is forbidden to be changed at random during

measurement.

4-7 Take caution against shock in the course of measuring voltage higher

than DC 60V & AC 30V.

4-8 Protection fuse should be replaced only with same type and same

specification.

4-9 After operation is finished, set function switch at OFF range to save

battery power.

4-10 If the meter is without usage for long time, take out battery to avoid

damage by battery leakage.

5. GENERAL SPECIFICATIONS

5-1 Max Voltage between input terminal and Earth Ground: CAT

600V

5-2 Over-range Indication: display “1” or “-1” for the significant digit.

5-3 Automatic display of negative polarity “-” .

5-4 Low Battery Indication: “ ” displayed.

5-5 Display: 3 1/2 digit LCD with a max. reading of 1999.

5-6 Manual range control

5-7 Auto Power Off: The meter will switch to standby mode when power on

after approx. 30 minutes. Press the

LIGHT

key or rotate the rotary

switch to

OFF

position, then power on to exit standby mode.

5-8 Fuse protection: 200mA/250V PPTC Resettable Fuse

5-9 Power supply: 9V battery (6F22 or NEDA 1604)

5-10 Operating Temp.: 0

to 40

(relative humidity <85%)

5-11 Storage Temp.:-10

to 50

((relative humidity <85%)

5-12 Guaranteed precision Temp.: 23

±

5

(relative humidity <70%)

5-13 Dimension: 150x106x36mm

5-14 Weight: approx. 250g (including battery)

6. Testing Specifications

Accuracy is specified for a period of year after calibration and at 18

to

28

(64

to 82

) with relative humidity to 70%.

6-1 DC Voltage

Range

Resolution

Accuracy

2V

1mV

±(0.5% of rdg + 2 digits)

20V

10mV

200V

100mV

600V

1V

±(0.8% of rdg + 2 digits)

-- Impedance: 1M

-- Overload protection: 600V DC or AC rms

6-2 AC Voltage

Range

Resolution

Accuracy

200V

100mV

±(1.5% of rdg + 5 digits)

600V

1V

-- Impedance: 450k

-- Overload protection: 600V DC or AC rms

-- Frequency Range: 40 to 400Hz

-- Response: average, calibrated in rms of sine wave

6-3 DC Current

Range

Resolution

Accuracy

2mA

1

μ

A

±(1.2% of rdg + 2 digits)

20mA

10

μ

A

200mA

100

μ

A

10A

10mA

±(2.0% of rdg + 3 digits)

-- Overload protection: 200mA/250V PPTC Resettable Fuse

Note: 10A range is not fused, 10A up to 10 seconds

6-4 Resistance

Range

Resolution

Accuracy

200

0.1

±(1.0% of rdg + 3 digits)

2k

1

±(1.0% of rdg + 2 digits)

20k

10

200k

100

2M

1k

±(1.5% of rdg + 3 digits)

-- Overload protection: 250V DC or AC rms

6-5 Temperature

Range

Accuracy

Resolution

-20~150

± ( 3

+ 1digit )

1

150~1000

± ( 3% of rdg + 2digits )

-- NiCr-NiSi K-type sensor

-- Overload protection: 200mA/250V PPTC Resettable Fuse

6-6 Battery test

Range

Accuracy

Load current

Resolution

±(5.0% of rdg + 5 digits)

Approx. 50mA

1mV

Approx. 10mA

10mV

-- Overload protection: 200mA/250V PPTC Resettable Fuse

6-7 Diode and Audible continuity test

Range

Description

Test Condition

Display read approximately

forward voltage of diode

Forward DC current

approx. 1mA

Reversed DC voltage

approx. 3V

Built-in buzzer sounds if

resistance is less than 50

Open circuit voltage

approx. 3V

Overload protection: 250V DC or AC rms