background image

770A

DIGITAL MULTIMETER

OPERATOR’S MANUAL

1.

Overview

The multimeter is characterized at slim size, portable, stable performance and

anti-dropping capacity. Using 3 1/2 digits LCD monitor with character 28mm high,

they offer clear readings.

With overall circuitry design centering on large-scale IC

A/D converters in conjunction and over-load protection circuit, the meters give

excellent performance and exquisite making as a handy utility instrument.

The meters can be used to measure DC & AC voltage, DC & AC current, resistance,

capacitor, transistor hFE, Non Contact AC Voltage (NCV) detection, positive diode

voltage fall and audible continuity.

2. Panel Layout

1

NCV detection area: Non Contact AC Voltage (NCV) detection area.

2

CDS sensor: The CDS sensor can reaction to the ambient brightness range,

then automatically control the LCD backlight to lighten or go out.

3

NCV red light: Non Contact AC Voltage (NCV) detection red light.

4

NCV green light: Non Contact AC Voltage (NCV) detection green light.

5

LCD display: 3 1/2 digits LCD display.

6

POWER key: Press the “POWER” key to power on or off the meter.

7

HOLD key: Press the “HOLD” key to lock display value, and the “

” sign will

appear on the display, press it again to exit.

8

NCV Key: Press and hold the “NCV” key, the meter enters Non Contact AC

Voltage (NCV) detection, the NCV green LED light will light up, free it to exit.

9

Rotary Switch: Use this switch to select functions and ranges.

10

mA

mA and capacitor Input Jack

11

20A: 20A Input Jack

12

V

Input Jack

13

COM: COM Input Jack

14

Transistor hFE test Input Jack

15

Crust of meter

16

Protective casing

3. Safety Information

3-1 The meter is designed according to IEC-1010 concerning electronic measuring

instruments with an over-voltage category 1000V (CAT

) and pollution 2.

3-2 Follow all safety and operating instructions to ensure that the meter is used

safely and is kept in good operating condition.

3-3 safety symbols:

Important safety information, refer to the operating manual.

Dangerous voltage may be presence.
Double insulation (protection Class II)

4. Special Cautions for Operation

4-1 The meters can be safe only according to standard procedures when used in

conjunctions with the supplied test leads. To replace damaged test leads with only

the same model or same electric specifications.

4-2 To avid risk of electric shock, do not use the meters before the cover is in place.

4-3 The range switch should be right position for the testing.

4-4 To avoid electric shock and damaging the instruments, the input signals are

forbidden to exceed the specified limits.

4-5 When measuring TV set or switched power, attention should be paid to the

possible pulses that may bring destruction to the circuit.

4-6 Range switch position is forbidden to be changed at random during

measurement.

4-7 Take caution against shock in the course of measuring voltage higher than DC

60V & AC 30V.

4-8 Protection fuse should be replaced only with same type and same specification.

4-9 After operation is finished, power OFF the meter to save battery power.

4-10 If the meter is without usage for long time, take out battery to avoid damage by

battery leakage.

5. GENERAL SPECIFICATIONS

5-1 Max Voltage between input terminal and Earth Ground: CAT

1000V

5-2 Over-range Indication: display “1” or “-1” for the significant digit.

5-3 Automatic display of negative polarity “-” .

5-4 Low Battery Indication: “ ” displayed.

5-5 Display: 3 1/2 digit LCD with a max. reading of 1999.

5-6 Manual range control

5-7 Auto Power Off: The meter will switch to standby mode when power on after

approx. 20 minutes. Press the “POWER” key two times to power on the meter

again.

5-8 Auto LCD backlight

5-9 Fuse protection: 200mA/500V & 20A/500V Fast Fuse

5-10 Power supply: 9V battery (6F22 or NEDA 1604)

5-11 Operating Temp. : 0

to 40

(relative humidity <85%)

5-12 Storage Temp. : -10

to 50

((relative humidity <85%)

5-13 Guaranteed precision Temp. : 23

±

5

(relative humidity <70%)

5-14 Dimension: 195x88x40mm

5-15 Weight: approx. 350g (including battery)

6. Testing Specifications

Accuracy is specified for a period of year after calibration and at 18

to 28

(64

to 82

) with relative humidity to 70%.

6-1 DC Voltage

Range

Resolution

Accuracy

200mV

0.1mV

±(0.5% of rdg + 2 digits)

2V

1mV

20V

10mV

200V

100mV

1000V

1V

±(0.8% of rdg + 2 digits)

-- Impedance: 10MΩ

-- Overload protection: 1000V DC or 750V AC rms

6-2 AC Voltage

Range

Resolution

Accuracy

2V

1mV

±(1.0% of rdg + 2 digits)

20V

10mV

200V

100mV

750V

1V

±(1.2% of rdg + 3 digits)

-- Impedance: 10MΩ (2V range is 1MΩ)

-- Overload protection: 1000V DC or 750 AC rms

-- Frequency Range: 40 to 400Hz

-- Response: average, calibrated in rms of sine wave

6-3 DC Current

Range

Resolution

Accuracy

2mA

1μA

±(1.2% of rdg + 2 digits)

20mA

10μA

200mA

100μA

20A

10mA

±(2.0% of rdg + 3 digits)

-- Overload protection: 200mA/500V & 20A/500V Fast Fuse

Note: 20A up to 10 seconds

6-4 AC Current

Range

Resolution

Accuracy

2mA

1μA

±(1.5% of rdg + 3 digits)

200mA

100μA

20A

10mA

±(2.5% of rdg + 5 digits)

-- Overload protection: 200mA/500V & 20A/500V Fast Fuse

Note: 20A up to 10 seconds

-- Frequency Range: 40 to 400Hz

-- Response: average, calibrated in rms of sine wave

6-5 Resistance

Range

Resolution

Accuracy

200Ω

0.1Ω

±(1.0% of rdg + 3 digits)

2kΩ

±(1.0% of rdg + 2 digits)

20kΩ

10Ω

200kΩ

100Ω

20MΩ

10kΩ

±(1.5% of rdg + 3 digits)

200MΩ

100kΩ

±(5.0% of rdg + 10 digits)

-- Overload protection: 500V DC or AC rms

6-6 Capacitance

Range

Accuracy

Resolution

2nF

±(2.5% of rdg + 5 digits)

1pF

20nF

10pF

200nF

100pF

2µF

1nF

20µF

10nF

200µF

±(5.0% of rdg + 10 digits)

100nF

-- Overload protection: 200mA/500V Fast Fuse

6-7 Diode and Audible continuity test

Range

Description

Test Condition

Display read approximately

forward voltage of diode

Forward DC current

approx. 1mA

Reversed DC voltage

approx. 3V

Built-in buzzer sounds if

resistance is less than 50Ω

Open circuit voltage

approx. 3V

Overload protection: 500V DC or AC rms

6-8 Transistor hFE test

Test range: 0-1000

Ib=10µA, Vce=3.0V Approx.

6-9 Non Contact AC Voltage (NCV) detection

Test voltage range: 90V

1000V AC rms

The NCV red LED light and green LED light will light up alternately together with

sound.

Отзывы: