Installation
i. Installation requirements
Space requirements
The drawing show the minimum space safety
requirements:
Fig. i.1
geodyna 8250
Fig. i.2
geodyna 8200
Each drawing has two sets of dimensions:
1
from the wall to the centre of the holes: on
the left and top of the drawing
2
from the wall to the outline of the cabinet: on
the right and bottom of the drawing
Floor requirements
The floor must be:
- horizontal; +/- 1° tolerance
- even; tolerance within 2 mm
- able to bear the weight of the balancer as stated in
Chapter 2 of the Operator’s Manual.
The floor on which the balancer will be installed should
not transmit vibrations from other devices or from
outside the building. External vibrations may affect the
accuracy of the unit.
Note:
The balancer should be positioned on the
floor directly. Do not use spacers to fill
gaps.
If the above conditions are satisfied, the balancer does
not need fixing to the floor.
Power supply requirements
Refer to Chapter 2 of the Operator’s Manual for power
mains requirements.
WARNING: ENSURE THAT AN APPROVED
WALL MAINS OUTLET IS AVAILABLE.
WARNING: NEVER GUIDE POWER SUPPLY
CABLES OVER THE FLOOR, UNLESS
PROTECTED BY AN APPROVED COVER.
....... ........