8.
End of test
(determined by
the application)
"Q:xxx"
The application stops the test. One of the
test
results
is shown. The user can check the other
test results by pushing the
Left
and
Right
keys of
the DADS Terminal.
Refer to End Condition of a Measurement
Recording to get to know when the test is ended
by the application.
9.
Confirm
Name of the
test or "No
tests"
Confirm by pressing the
Enter
key of the DADS
Terminal, or by clicking
Cancel
button on the main
screen.
Test list measurement procedure
Test list editor to generate a test or a test list. A test list is a certain number of measurement
definitions, identified by a test name. Refer to the
information about it.
Step
Action
Terminal LCD
Description
1.
Define test(s)
Name of the
test or "No
tests"
Define one or more tests with the
test list editor
.
information on defining tests.
2.
Choose test
list group
Name of a test
from the test
list
Choose one of the test list groups (by using the
group selection list in the mainframe).
3.
Choose test
Name of a test
from the test
list
Choose the test that you want to run by using the
Up
and
Down
keys of the DADS Terminal.
4.
Request
measurement
start
Name of the
test to run
Request the measurement start by pushing the
Enter
key on the DADS Terminal.
5.
Start the motor
"Start motor"
Start the motor
(pushing the green lower button
on the DADS Terminal).
In case that the user pushes the
Left
key of the
DADS Terminal the test is cancelled and the
application turns back into the idle state. Start from
step 1 to redo a test.
6.
Measurement
running
"Test"
The measurement has started and the application
displays the graph torque (mNm) vs. liquid quantity
(ml/100g).
Wait for the end of test
.
Refer to Start Condition of a Measurement
Recording to get to know when the test is started
by the application.
The test can be aborted by stopping the motor.