Page 62/81
Default test procedure
Follow these steps to do a default test procedure. The test list has to be empty; otherwise a
test from the test list is taken.
Step
Action
Terminal LCD
Description
1.
Request
measurement
start
"No test list"
Request the measurement start by pushing the
Enter
key on the DADS Terminal.
2.
Acknowledge
"Default test?"
Acknowledge by pressing the
Enter
key on the
3.
Enter carbon
black type
"Class: Hard"
Choose the carbon black type (hard or soft) by
using the
Up, Down, and Enter
keys on the
4.
Enter test
method
"Method:
Normal"
Choose the test method (normal or compressed)
by using the
Up, Down, and Enter
keys on the
5.
Enter weight
"Weight:
xx.xx g"
Change the weight by using the
arrow keys
of
the DADS terminal. Terminate by pressing
Enter
on the DADS Terminal.
6.
Start the motor
"Start motor"
Start the motor
, by pushing the motor start
button of the Absorptometer (on DABS Front
door of the DABS terminal).
In case that the user pushes the
Left
key of the
DADS Terminal the test is cancelled and the
application turns back into the idle state. Start
from step 1 to redo a test.
7.
Measurement
is running
"Date"
The test name is automatically generated as
date & time and is used to file the data. The
measurement has started and the application
displays the graph torque (mNm) vs. liquid
quantity (ml/100g).
Wait for the end of test
.
Refer to Start Condition of a Measurement
Recording to get to know when the test is started
by the application.
The test can be aborted by stopping the motor.
Refer to Abort a Test for more information.
8.
End of test
(determined by
the application)
"Q:xxx"
The application stops the test. One of the
test
results
is shown. The user can check the other
test results by pushing the
Left
and
Right
keys
of the DADS terminal.
Refer to End Condition of a Measurement
Recording to get to know when the test is ended