Note:
The user can accept a TLS test by clicking on the
Accept
button or by pushing two
times the
Enter
button on the DADS terminal.
A TLS test will be rejected in case the clicks the
Reject
button, or push the
Enter
on
keyboard, or pushing the
Left
button on the DADS terminal.
Attenuation
This function is only used, and available, with the Absorptometer machines of type B
The attenuation test facility is provided for the type B machines, to test the
full scale
recovery
of the torque sensing system. This one should have a value of approximately
3 s
,
according to the ASTM D2414 specifications. The test should be repeated until the machine
B fit's to the specifications.
At first a window with the last value and the test date is shown.