6-37
6.4 Warning conditions (printing continuously performed)
6.4.1 Excitation V-ref. Review, Excitation V-ref. Review 2
6.4.2 Viscosity Reading Instability
Viscosity Reading
Instability
Resume the
operation.
Is it normally measured?
Yes
No
Perform viscometer
calibration
Perform the viscometer
operation test on the
Circulation system control
screen.
Replace the viscometer filter.
Disassemble and clean the
viscometer.
Replace the viscometer
Is it normally measured?
Is it normally measured?
Yes
Yes
No
No
Viscosity Reading
Instability
during installation
(calibration)
Re-execute viscometer
calibration (2-3 times).
Does calibration finish
normally?
No
Yes
Press [OK], then the
coefficient is determined
Excitation V-ref. Review
Resume the operation.
Readjust the excitation setup.
Содержание IJ RX2
Страница 1: ...Service Manual HITACHI Printer Model RX2 Revision Feb 2014 Version A ...
Страница 13: ...1 8 2 External views Rear side 1 2 2 Main body internal ...
Страница 14: ...1 9 1 2 3 Print head ...
Страница 101: ...3 26 Circuit diagram of EZJ127A ...
Страница 102: ...3 27 Circuit diagram of EZJ127B ...
Страница 116: ...3 41 Circuit diagram of EZJ129 ...
Страница 164: ...4 40 5 Reset the time of the R air filter to 0 on the Parts usage time management screen ...
Страница 247: ...7 Attached Drawing 7 1 Circulation System Diagram Circulation System Diagram 7 1 ...
Страница 248: ...7 2 Electrical Connection Diagram 7 2 1 Electrical Connection Diagram RX2 S Standard model 7 2 ...
Страница 249: ...7 3 7 2 2 Electrical Connection Diagram RX2 B Basic model 7 3 ...
Страница 252: ... Nozzle diameter 65μm 7 6 7 4 Dimensions around charge electrode and deflection electrode 循環系統図 ...