80
Communication Method
Standard Event Status Register (SESR)
The Standard Event Status Register is an 8-bit register. If any bit in the Standard Event Status
Register is set to “1” (within the enabled bits set by the Standard Event Status Enable Resister),
bit 5 (ESB) of the Status Byte Register is set to “1”.
See: "Standard Event Status Register (SESR) and Standard Event Status Enable Register
The Standard Event Register is cleared in the following situations:
•
When a
*CLS
command is executed
•
When an event register query (
*ESR?
) is executed
•
When the power is turned on again
Bit 7
PON
Power-On Flag
Set to “1” when the power is turned on, or upon recovery from an outage.
Bit 6
URQ
(Unused)
Not used by this device
User Request
Bit 5
CME
Command error (The command to the message terminator is ignored.)
Set to “1” when a received command contains a syntactic or semantic error.
•
Program header error
•
Incorrect number of data parameters
•
Invalid parameter format
•
Received a command not supported by the device
Bit 4
EXE
Execution error
Set to “1” when a received command cannot be executed for some reason.
•
The specified data value is outside of the set range
•
The specified setting data cannot be set
•
Execution is prevented by some other operation being performed
Bit 3
DDE
Device-dependent error
Set to “1” in the event of a backup error, self-test error, communications
parameter error, communications timeout, or other issue.
Bit 2
QYE
Query error (The output queue is cleared.)
Set to “1” when a query error is detected by the output queue control.
•
When the data overflows the output queue
•
When data in the output queue has been lost
Bit 1
RQC
(Unused)
Not used by this device.
Request control
Bit 0
OPC
Operation complete
•Execution of an
*
OPC
command.
•
Completion of operations of all messages up to the
*
OPC
command.
HIOKI SW1001A961-04
Содержание SW9001
Страница 2: ...HIOKI SW1001A961 04...
Страница 42: ...38 Channel Delay Function HIOKI SW1001A961 04...
Страница 48: ...44 Scan Measurement Example HIOKI SW1001A961 04...
Страница 50: ...46 Checking the Device Status HIOKI SW1001A961 04...
Страница 52: ...48 Initialization Settings HIOKI SW1001A961 04...
Страница 118: ...114 Sample Programs 3 Select FILE Save All HIOKI SW1001A961 04...
Страница 160: ...156 Outline Drawings 12 4 Outline Drawings SW1001 Unit mm Tolerance 0 2 HIOKI SW1001A961 04...
Страница 161: ...157 Outline Drawings SW1002 Unit mm Tolerance 0 2 12 Appendix HIOKI SW1001A961 04...
Страница 162: ...158 Outline Drawings HIOKI SW1001A961 04...
Страница 164: ...160 HIOKI SW1001A961 04...
Страница 167: ...HIOKI SW1001A961 04...
Страница 168: ...HIOKI SW1001A961 04...
Страница 169: ...HIOKI SW1001A961 04...