76
Communication Method
Output queue and input buffer
Output queue
Response messages are stored in the output queue until read by the controller. The output queue
is also cleared in the following circumstances.
•
When the power is turned on
• Query error
At a minimum, the device’s output queue can buffer a 64-byte response message.
When using the RS-232C interface, data in excess of that length may result in the output buffer
being cleared due to a query error.
When using the LAN or USB interface, query operation will enter a wait state once the buffer is full
and continue in that state until the data is received.
Input buffer
The input buffer can accommodate a minimum of 256 characters.
When using the RS-232C interface, data may be lost in the event of a buffer overrun.
When using the LAN or USB interface, operation will enter a wait state once the buffer becomes full
and continue in that state until the buffer is empty.
HIOKI SW1001A961-04
Содержание SW9001
Страница 2: ...HIOKI SW1001A961 04...
Страница 42: ...38 Channel Delay Function HIOKI SW1001A961 04...
Страница 48: ...44 Scan Measurement Example HIOKI SW1001A961 04...
Страница 50: ...46 Checking the Device Status HIOKI SW1001A961 04...
Страница 52: ...48 Initialization Settings HIOKI SW1001A961 04...
Страница 118: ...114 Sample Programs 3 Select FILE Save All HIOKI SW1001A961 04...
Страница 160: ...156 Outline Drawings 12 4 Outline Drawings SW1001 Unit mm Tolerance 0 2 HIOKI SW1001A961 04...
Страница 161: ...157 Outline Drawings SW1002 Unit mm Tolerance 0 2 12 Appendix HIOKI SW1001A961 04...
Страница 162: ...158 Outline Drawings HIOKI SW1001A961 04...
Страница 164: ...160 HIOKI SW1001A961 04...
Страница 167: ...HIOKI SW1001A961 04...
Страница 168: ...HIOKI SW1001A961 04...
Страница 169: ...HIOKI SW1001A961 04...