![Hioki SW9001 Скачать руководство пользователя страница 74](http://html1.mh-extra.com/html/hioki/sw9001/sw9001_instruction-manual_2137434074.webp)
70
Communication Command Forwarding Function
Connection method
When using the communication command forwarding function, connect the
INSTRUMENT
RS-232C
connector of the device and the
RS-232C
connector of the instrument.
Rear
SW1001 (SW1002)
Instrument
Rear
Use the RS-232C
cross cable.
INSTRUMENT
RS-232C
connector
RS-232C
connector
Set the communication speed of the instrument based on the communication speed setting of the
device.
Setting the transmission speed
Set the commands shown below for the transmission speed of the communication command
forwarding function.
:SYSTem:COMMunicate:FORWard:RS232C:SPEED <9600/19200/38400>
The speed can be selected from 9600 bps, 19200 bps, or 38400 bps.
The speed setting is immediately changed.
Other RS-232C communication parameters are fixed to 8 for data bit length, 1 for stop bit, None for
parity check, and None for flow control.
See: "2.6 Setting the Communication Setting Mode" (p. 22)
Setting the receive timeout
You can set the timeout time to use when receiving data from the instrument by means of the
communication command forwarding function.
Sending a query to the instrument will result in a timeout error if the set timeout time elapses before
a response is received from the instrument.
Timeout can be set by the command below.
:SYSTem:COMMunicate:FORWard:TIMeout <
Time
>
Time: Timeout time (s)
HIOKI SW1001A961-04
Содержание SW9001
Страница 2: ...HIOKI SW1001A961 04...
Страница 42: ...38 Channel Delay Function HIOKI SW1001A961 04...
Страница 48: ...44 Scan Measurement Example HIOKI SW1001A961 04...
Страница 50: ...46 Checking the Device Status HIOKI SW1001A961 04...
Страница 52: ...48 Initialization Settings HIOKI SW1001A961 04...
Страница 118: ...114 Sample Programs 3 Select FILE Save All HIOKI SW1001A961 04...
Страница 160: ...156 Outline Drawings 12 4 Outline Drawings SW1001 Unit mm Tolerance 0 2 HIOKI SW1001A961 04...
Страница 161: ...157 Outline Drawings SW1002 Unit mm Tolerance 0 2 12 Appendix HIOKI SW1001A961 04...
Страница 162: ...158 Outline Drawings HIOKI SW1001A961 04...
Страница 164: ...160 HIOKI SW1001A961 04...
Страница 167: ...HIOKI SW1001A961 04...
Страница 168: ...HIOKI SW1001A961 04...
Страница 169: ...HIOKI SW1001A961 04...