18
Connecting the Measurement Cable
2.4 Connecting the Measurement Cable
WARNING
Connect the measurement cable with the measuring object disconnected from
the measurement cable. The measuring object may be shorted depending on the
condition of the module switching circuit.
Please provide a suitable measurement cable.
See: “12.1 Measurement cable” (p. 145)
Measuring object
Tip of the probe
Disconnect
Disconnect
1
Turn off the device.
2
Disconnect the probe from the
measuring object.
Front
3
4
3
Connect the measurement
cable connector to the module’s
connector.
4
Secure the measurement cable
connector using the screws.
For information on connecting the measuring object, see the instruction manual of each instrument.
HIOKI SW1001A961-04
Содержание SW9001
Страница 2: ...HIOKI SW1001A961 04...
Страница 42: ...38 Channel Delay Function HIOKI SW1001A961 04...
Страница 48: ...44 Scan Measurement Example HIOKI SW1001A961 04...
Страница 50: ...46 Checking the Device Status HIOKI SW1001A961 04...
Страница 52: ...48 Initialization Settings HIOKI SW1001A961 04...
Страница 118: ...114 Sample Programs 3 Select FILE Save All HIOKI SW1001A961 04...
Страница 160: ...156 Outline Drawings 12 4 Outline Drawings SW1001 Unit mm Tolerance 0 2 HIOKI SW1001A961 04...
Страница 161: ...157 Outline Drawings SW1002 Unit mm Tolerance 0 2 12 Appendix HIOKI SW1001A961 04...
Страница 162: ...158 Outline Drawings HIOKI SW1001A961 04...
Страница 164: ...160 HIOKI SW1001A961 04...
Страница 167: ...HIOKI SW1001A961 04...
Страница 168: ...HIOKI SW1001A961 04...
Страница 169: ...HIOKI SW1001A961 04...