![Hioki SW9001 Скачать руководство пользователя страница 147](http://html1.mh-extra.com/html/hioki/sw9001/sw9001_instruction-manual_2137434147.webp)
143
Accuracy Calculation Example
11.3
Accuracy Calculation Example
This section describes how to calculate the combined accuracy of the device and a connected
measuring instrument.
This process consists of calculating the measurement influence for each measuring instrument
(target instrument) used in the combination.
The measurement accuracy when using the device and a module is calculated using the following
formula:
Combination measurement accuracy = Instrument ac combination measurement influence
For more information about a given module’s measurement influences, see the module’s
specifications.
• SW9001
"Measurement influence" (p. 131)
•
SW9002
"Measurement influence" (p. 139)
Below shows an example of accuracy calculation.
Example 1: Use of the device in combination with the BT3562A
Used module
SW9001
Instrument
BT3562A
Ω
V function, 3 m
Ω
range, 6 V range, sampling SLOW
Measurement value
1 m
Ω
, 4 V
Resistance measurement accuracy = BT3562A resistance measurement ac SW9001
measurement influence (BT3562A 3 m
Ω
range)
= (±0.5% rdg. ± 10 dgt.) + (±0.1% f.s.)
= (±0.5% × 1 m
Ω
+ 0.0010 m
Ω
) + [±(0.1% × 3.1000 m
Ω
)]
= (±0.0060 m
Ω
) + (±0.0031 m
Ω
)
= ±0.0091 m
Ω
(total error)
Voltage measurement accuracy = BT3562A voltage measurement ac SW9001
measurement influence (all BT3562A voltage ranges)
= (±0.01% rdg. ± 3 dgt.) + (±5 µV)
= (±0.01% × 4 V ± 0.00003 V) + (±5 µV)
= (±0.00043 V) + (±5 µV)
= ±0.000435 V (total error)
11
Multiplexer Module
HIOKI SW1001A961-04
Содержание SW9001
Страница 2: ...HIOKI SW1001A961 04...
Страница 42: ...38 Channel Delay Function HIOKI SW1001A961 04...
Страница 48: ...44 Scan Measurement Example HIOKI SW1001A961 04...
Страница 50: ...46 Checking the Device Status HIOKI SW1001A961 04...
Страница 52: ...48 Initialization Settings HIOKI SW1001A961 04...
Страница 118: ...114 Sample Programs 3 Select FILE Save All HIOKI SW1001A961 04...
Страница 160: ...156 Outline Drawings 12 4 Outline Drawings SW1001 Unit mm Tolerance 0 2 HIOKI SW1001A961 04...
Страница 161: ...157 Outline Drawings SW1002 Unit mm Tolerance 0 2 12 Appendix HIOKI SW1001A961 04...
Страница 162: ...158 Outline Drawings HIOKI SW1001A961 04...
Страница 164: ...160 HIOKI SW1001A961 04...
Страница 167: ...HIOKI SW1001A961 04...
Страница 168: ...HIOKI SW1001A961 04...
Страница 169: ...HIOKI SW1001A961 04...