106
Message Reference
Query relay opening/closing frequency
Syntax
Query
:SYSTem:MODule:COUNt? <
Slot No.
>
[,<Relay No.>]
Response
<Slot No.> = 1 to 3 (NR1) (SW1001)
1 to 12 (NR1) (SW1002)
<Opening/closing frequency> = 0 to 1000000000 (NR1)
Description
Returns to the specified relay opening/closing frequency.
Example
:SYSTem:MODule:COUNt? 1
101257
Note
•
The relay opening/closing frequency is saved in the non-volatile memory inside each
module. If no relay number is specified, the query will return the count for the relay for
the selected slot that has been turned off and on the most times. (See: p. 136, p. 142)
•
Specifying a slot without a module will result in an execution error.
(11) Initialization
Initialize the device
Syntax
Command
:SYSTem:PRESet
:STATus:PRESet
Description
Resets the device to its initial state.
See: "6.1 Initialization Settings" (p. 47)
The operation is the same for
:SYSTem:PRESet
and
:STATus:PRESet
.
Example
:SYST:PRES
Note
•
The communication conditions are not initialized.
•All closed channels are opened.
•
Initialized settings are not preserved when the power is turned off. To save initialized
settings, execute the
SYSTem:BACKup
command.
See: "Backing up settings" (p. 105)
(12) Test
Relay contact welding test
Syntax
Command
:TEST:RELayshort <
Slot No.
>,<
Test item
>
<Slot No.> = 1 to 3 (NR1) (SW1001)
1 to 12 (NR1) (SW1002)
<Test item> = CH/BUS/OPEN
CH:
Channel relay test
BUS:
Bus relay test
OPEN:
Open the relay after all test is finished.
Description
Execute relay contact welding test to each slot.
The test must be executed to both channel relay and bus relay.
For details, see "Checking for relay contact welding" (p. 26)
Example
:TEST:RELayshort 1,CH
Execute the test to channel relay of
SLOT 1
.
HIOKI SW1001A961-04
Содержание SW9001
Страница 2: ...HIOKI SW1001A961 04...
Страница 42: ...38 Channel Delay Function HIOKI SW1001A961 04...
Страница 48: ...44 Scan Measurement Example HIOKI SW1001A961 04...
Страница 50: ...46 Checking the Device Status HIOKI SW1001A961 04...
Страница 52: ...48 Initialization Settings HIOKI SW1001A961 04...
Страница 118: ...114 Sample Programs 3 Select FILE Save All HIOKI SW1001A961 04...
Страница 160: ...156 Outline Drawings 12 4 Outline Drawings SW1001 Unit mm Tolerance 0 2 HIOKI SW1001A961 04...
Страница 161: ...157 Outline Drawings SW1002 Unit mm Tolerance 0 2 12 Appendix HIOKI SW1001A961 04...
Страница 162: ...158 Outline Drawings HIOKI SW1001A961 04...
Страница 164: ...160 HIOKI SW1001A961 04...
Страница 167: ...HIOKI SW1001A961 04...
Страница 168: ...HIOKI SW1001A961 04...
Страница 169: ...HIOKI SW1001A961 04...