Internal Circuit Configuration
PNP setting
Use ISO_COM for the common terminal for both the input and output signals.
PLC, etc.
Output
Common
Do not connect external power to one pin.
This device
3
ISO_5 V
1
SCAN
6
SCAN_RESET
4
CLOSE
8
ISO_COM
PLC, etc.
Input
Common
2 k
Ω
1 k
Ω
10
Ω
Zener voltage 30 V
Internal isolated common
(It is isolated from the ground for the device.)
PNP
EXT. I/O MODE
selector
Internal isolated
power supply
5 V
30 V max.
Electrical specifications
Input signal
Input format
Isolated non-voltage contact input using a photocoupler
(Compatible with sinking/sourcing current output)
Input ON
Residual voltage 1 V (input ON current 4 mA (reference value))
Input OFF
OPEN (interrupting current 100
µ
A or less)
Output signal
Output format
Isolated open drain output using a photocoupler (non-polar)
Maximum load voltage
30 V DC max.
Maximum output current
50 mA/channel
Residual voltage
1 V or less (load current 50 mA)/0.5 V or less (load current 10
mA)
Internal
isolated power
supply
Output voltage
Sinking output compatible: +4.2 V to +5.8 V
Sourcing output compatible:
-
4.2 V to
-
5.8 V
Maximum output current
100 mA
External power supply
input
None
Insulation
Floating from the protective grounded potential and the
measurement circuit
Insulation rating
Voltage between the module and ground: 50 V DC, 30 V AC
rms, 42.4 V AC peak or less
Содержание SW1001
Страница 2: ......
Страница 42: ...38 Precautions for Measurement...
Страница 48: ...44 Resetting Scan Operation...
Страница 50: ...46 Checking the Device Status...
Страница 52: ...48 Initialization Settings...
Страница 118: ...114 Sample Programs 3 Select File Save All 4 Check the save location and click Save as...
Страница 156: ...152 Outline Drawings 12 4 Outline Drawings SW1001 Unit mm Tolerance 0 2...
Страница 157: ...153 Outline Drawings SW1002 Unit mm Tolerance 0 2 12 Appendix...
Страница 158: ...154 Outline Drawings...
Страница 161: ...16 01 EN...
Страница 162: ......
Страница 163: ......
Страница 164: ......