Precautions for Measurement
Influence of thermoelectric power
Be aware of the thermoelectric power when a voltage accuracy of a few µV is required.
Install the device and instrument in a constant temperature environment and fully adjust them to the
ambient temperature before use.
Especially prevent uneven temperatures at the terminal.
When creating a measurement cable, use a connector and terminal of a material with low
thermoelectric voltage, such as brass (nickel plate), pure copper (+ gold plate), etc.
Exercise care to keep the contact surface clean.
High resistance and minute current measurement
In a high humidity environment, high resistance or minute current measurement may be affected by
leak current.
Do not touch the module board with your bare hand. Grease can cause leak current, which may
affect high resistance and minute current measurement.
Do not bundle the measurement cables when measuring minute current. Doing so may cause leak
current due to the capacity of the cables, resulting in an error in the measured value.
Caution regarding noise
Do not bundle the measurement cables or connection cables for the instrument and the power line.
Noise on the power line may cause malfunction of channel switching or an error in the measured
value.
When the measured value cannot be stabilized due to noise, connecting the ground (
GND
terminal)
of the device and the ground of the instrument may improve the condition.
Caution regarding load
When a load is applied to the measurement cable, the contact area of the connector becomes
unstable and the contact resistance is increased, which can cause the measured value to be
unstable.
When attaching the module to the main frame, be sure to secure the panel using screws. If the
panel cannot support the load when a vertical load is applied to the main frame, the module board
is stressed and may malfunction.
Содержание SW1001
Страница 2: ......
Страница 42: ...38 Precautions for Measurement...
Страница 48: ...44 Resetting Scan Operation...
Страница 50: ...46 Checking the Device Status...
Страница 52: ...48 Initialization Settings...
Страница 118: ...114 Sample Programs 3 Select File Save All 4 Check the save location and click Save as...
Страница 156: ...152 Outline Drawings 12 4 Outline Drawings SW1001 Unit mm Tolerance 0 2...
Страница 157: ...153 Outline Drawings SW1002 Unit mm Tolerance 0 2 12 Appendix...
Страница 158: ...154 Outline Drawings...
Страница 161: ...16 01 EN...
Страница 162: ......
Страница 163: ......
Страница 164: ......