120
SW1001, SW1002 Input Specifications/Output Specifications/Measurement Specifications
9.2
SW1001, SW1002 Input Specifications/Output
Specifications/Measurement Specifications
Basic specifications
Number of slots
3 slots (SW1001), 12 slots (SW1002)
Analog bus
Number of systems
2-wire 4-system (automatic selection based on the module
and connection method)
Terminal
BNC receptacle × 4 (4-terminal pair)
Banana terminal receptacle × 4 (4-wire)
Banana terminal receptacle × 2 (2-wire)
Maximum input voltage
60 V DC, 30 V AC rms, 42.4 V peak
Maximum rated voltage
to earth
60 V DC
(Anticipated transient overvoltage: 60 V)
Compatible module
Model SW9001 Multiplexer Module
Model SW9002 Multiplexer Module
Number of instruments
to be connected
Maximum 2
One 2-wire + one 4-wire or
one 2-wire + one 4-terminal pair
Содержание SW1001
Страница 2: ......
Страница 42: ...38 Precautions for Measurement...
Страница 48: ...44 Resetting Scan Operation...
Страница 50: ...46 Checking the Device Status...
Страница 52: ...48 Initialization Settings...
Страница 118: ...114 Sample Programs 3 Select File Save All 4 Check the save location and click Save as...
Страница 156: ...152 Outline Drawings 12 4 Outline Drawings SW1001 Unit mm Tolerance 0 2...
Страница 157: ...153 Outline Drawings SW1002 Unit mm Tolerance 0 2 12 Appendix...
Страница 158: ...154 Outline Drawings...
Страница 161: ...16 01 EN...
Страница 162: ......
Страница 163: ......
Страница 164: ......