99
Message Reference
(5) Scan Function
Register and query scan list
Syntax
Command
[:ROUTe]:SCAN<(@
Channel list
)>
Query
[:ROUTe]:SCAN?
Response
<(@Channel list)>
Specify the channel list in the following formats.
<CH>,<CH>, ..., <CH> Specify the channels to be measured individually.
<CHm>:<CHn> Specify from CHm to CHn consecutively.
Description
Registers the scan function scan list (channel list to be switched). Specify the channel
using the combined number of the slot number and channel number inside the module
<Slot No.> × 100 + channel number inside the module).
Up to 1000 channels can be registered for the scan list.
(@) can be omitted.
Example
:SCAN (@101,102,103,201,202)
Scanning is performed in order of 101, 102, 103, 201, and then 202.
:SCAN (@101:312)
Scanning is performed from 101 to 312.
:SCAN 1001:1222
Scanning is performed from 1001 to 1222.
:SCAN?
(@101,102,103,201,202)
Scanning is set to be performed in order of 101, 102, 103, 201, and then 202.
Note
An execution error occurs during scan operation (command).
Adding to the scan list
Syntax
Command
[:ROUTe]:SCAN:ADD <(@
Channel list
)>
<(@Channel list)>
Specify the channel list in the following formats.
<CH>,<CH>,...,<CH>
Specify the channels to be measured individually.
<CHm>:<CHn>
Specify from CHm to CHn consecutively.
Description
Adds a list of channels to the end of the present scan list.
A total of up to 1000 channels may be registered. Attempting to add 200 channels if there
are presently 900 channels registered will result in an error, leaving the original scan list
of 900 channels (none of the 200 channels will be added). You can check the number of
channels that can be added by using the
:SCAN:SIZE?
query.
Example
:SCAN 101,102
:SCAN:ADD 201,202
Scan list is 101, 102, 201, and 202.
Note
An execution error occurs during scan operation (command).
Delete scan list
Syntax
Command
[:ROUTe]:SCAN:REMove
Description
Deletes the scan list.
Example
:SCAN:REM
The scan list is deleted.
Note
An execution error occurs during scan operation (command).
8
Communication Function
Содержание SW1001
Страница 2: ......
Страница 42: ...38 Precautions for Measurement...
Страница 48: ...44 Resetting Scan Operation...
Страница 50: ...46 Checking the Device Status...
Страница 52: ...48 Initialization Settings...
Страница 118: ...114 Sample Programs 3 Select File Save All 4 Check the save location and click Save as...
Страница 156: ...152 Outline Drawings 12 4 Outline Drawings SW1001 Unit mm Tolerance 0 2...
Страница 157: ...153 Outline Drawings SW1002 Unit mm Tolerance 0 2 12 Appendix...
Страница 158: ...154 Outline Drawings...
Страница 161: ...16 01 EN...
Страница 162: ......
Страница 163: ......
Страница 164: ......