106
Contact Check
Setting the judgment threshold value
The threshold value of a contact check judgment is set as an electrostatic capacitance value for
this instrument.
1
1
1
Press
[BASIC]
>
[C.CHECK]
on the setting
screen.
2
2
2
Change the value using
[▲]
or
[▼]
, and then
press
[ENTER]
.
1.0 nF
*
1
to
100.0 nF
*1. Default setting
Executing a contact check
When the contact check function is enabled, a contact check judgment is executed at the end of a
withstand voltage test or insulation resistance test.
See “12.3 Timing Charts” (p. 166).
Acquring the compensation value
When the contact check function is set to
[ON]
, execute open compensation before testing.
Acquire the compensation value in an open state, without a DUT connected, by the same operation
as at the usual start of a test.
1
Press
[ADJ]
on the measurement screen to start
a test on the compensation execution screen.
HIOKI ST5680A961-00