51
During Measurement
4.2 During Measurement
The instrument applies test voltage to equipment to be measured during
measurement.
1
When the test duration is set to
[ON]
1. When the
START
key is pressed, test voltage is applied to the equipment to be
measured, and the timer countdown starts. The remaining time is displayed in the test
duration indicator (TIMER).
2. During the response time,
[D]
is displayed on the left edge of the countdown time
indicator, and the measured value indicator becomes blank (no display). The display
during the response time is the same for both response time settings,
[AUTO]
and
[MANUAL]
.
3. After the response time has elapsed, the resistance measurement is performed for the
time specified for the timer, and the measured value is displayed. (Unit: M
Ω
) When the
comparator function is set, judgment is performed according to the test mode setting.
2
When the test duration is set to
[OFF]
1. When the
START
key is pressed, test voltage is applied to the equipment to be
measured, and the timer count-up starts. The time elapsed since the
START
key was
pressed is displayed in the test duration indicator (TIMER).
2. During the response time,
[D]
is displayed on the left edge of the count-up time
indicator, and the measured value indicator becomes blank (no display). The display
during the response time is the same for both response time settings,
[AUTO]
and
[MANUAL]
.
3. After the response time has elapsed, the resistance measurement is performed until
the
STOP
key is pressed, and the measured value is displayed. (Unit: M
Ω
) The
elapsed time is counted and displayed up to 999.9 s, but the resistance measurement
is performed until the
STOP
key is pressed. When the comparator function is set,
judgment is performed according to the test mode setting.
3
PASS/FAIL judgment display for the comparator function
Upper limit value ≤ Measured value
U.FAIL turns on.
Lower limit value < Measured value < Upper limit value
PASS turns on.
Measured value ≤ Lower limit value
L.FAIL turns on.
Refer to “3.7 Announcing the Judgment Results Using a Beep Sound” (p. 47) to set
beep sound for PASS/FAIL judgment.
When the upper and lower limit values are not set, the indicator displays “----” and
PASS/FAIL judgment is not performed. If either of the upper and lower limit values is set,
PASS/FAIL judgment is performed according to the value. (p. 44)
If the upper and lower limit values are set to values outside of the accuracy range, the
U.FAIL and L.FAIL lamps turn on at the same time.
4
The timer indicator indicates the test duration countdown. When the test
duration is not set, the indicator displays elapsed time.
10
9
8
7
6
5
4
3
2
1
Appx.
Ind.
Содержание ST5520
Страница 2: ......
Страница 8: ...vi Contents ...
Страница 22: ...14 Usage Notes ...
Страница 56: ...48 Announcing the Judgment Results Using a Beep Sound ...
Страница 82: ...74 Initializing the System Reset ...
Страница 88: ...80 Deleting the Panel Data ...
Страница 158: ...150 Command Reference ...
Страница 193: ...Appx 13 Rack Mounting Screw location dimensional drawing 10 9 8 7 6 5 4 3 2 1 Appx Ind ...
Страница 194: ...Appx 14 Dimensional Diagram Appx 8 Dimensional Diagram 166 2 215 1 14 1 80 1 ...
Страница 197: ...13 09 ...
Страница 198: ......
Страница 199: ......
Страница 200: ......
Страница 201: ......