155
Test duration
Definition of test
duration
Test duration = Response time + Measurement time
Setting range
Setting range
0.045 s (45 ms) to 999.999 s
Test duration
ON/OFF (- -)
Display
000.000 s setting
Screen display: 000.0 s, 0.000 s if less than 100 ms
“000.0 s” is displayed when OFF is set, “HOLD” is displayed at the end
of a test.
Operation
The test duration counts down from the specified time after a test
starts.
When OFF is set, the test duration counts up after a test starts.
“999.9 s” is displayed when the test duration exceeds the display
range.
Guaranteed
accuracy range
0.050 s (50 ms) to 999.999 s
Set resolution
1 ms
Test duration
accuracy
±0.005 s (setting range 0.045 s to 0.099 s)
±0.05 s (setting range 0.100 s to 9.999 s)
±0.5 s (setting range 10.000 s to 999.999 s)
Voltage (V)
Set voltage
Response time
Charge time
Measurement time
Discharge time
Test duration
Time (t)
FAST
SLOW
Time diagram
(FAST, SLOW: TEST signal OFF timing setting)
10
9
8
7
6
5
4
3
2
1
Appx.
Ind.
Содержание ST5520
Страница 2: ......
Страница 8: ...vi Contents ...
Страница 22: ...14 Usage Notes ...
Страница 56: ...48 Announcing the Judgment Results Using a Beep Sound ...
Страница 82: ...74 Initializing the System Reset ...
Страница 88: ...80 Deleting the Panel Data ...
Страница 158: ...150 Command Reference ...
Страница 193: ...Appx 13 Rack Mounting Screw location dimensional drawing 10 9 8 7 6 5 4 3 2 1 Appx Ind ...
Страница 194: ...Appx 14 Dimensional Diagram Appx 8 Dimensional Diagram 166 2 215 1 14 1 80 1 ...
Страница 197: ...13 09 ...
Страница 198: ......
Страница 199: ......
Страница 200: ......
Страница 201: ......