89
Checking Test Start and Test Results
7.2
Checking Test Start and Test Results
Press the instrument
START
button to implement the test. (You can start the test from either
communications or external control terminal as well.)
When a test finishes, the test results are output to the screen, communications, and external control
terminal (EXT. I/O).
IMPORTANT
If any of the judgment results are FAIL, it views that dielectric breakdown as having started and
the test
ends at that point in time.
If the judgment results are all PASS, implement testing until max. voltage is reached.
PASS judgment example
Total judgment results
Judgment results
*1
Present applied voltage value
FAIL judgment example (discharge FAIL at 2000 V)
Total judgment results
Judgment results
*1
Voltage value during
breakdown judgment
*1:
LCRC: XX
σ
Max. deviation of LC and RC values
DCHG: XX
σ
Max. deviation of discharge amount
AREA: XX
σ
Max. deviation of waveform surface area value
Vpeak XX%
Max. misalignment width from standard of the peak voltage value
FREQ XX%
Max. misalignment value from standard of the vibration frequency
Note: The values displayed are the judgment values and results of the present voltage values
7
Break down voltage test (BDV)
Содержание ST4030
Страница 2: ......
Страница 14: ...6 Usage Notes ...
Страница 32: ...24 Inspecting before Measurement and Verifying Operations ...
Страница 42: ...34 Table Copying ...
Страница 108: ...100 Setting Dielectric Break down Judgment Conditions ...
Страница 132: ...124 Terminal Open Error Setting ...
Страница 198: ...190 Editing Files and Folders ...
Страница 214: ...206 Interface Specifications ...
Страница 237: ...229 Dimensions Diagram 15 4 Dimensions Diagram Unit mm 215 1 348 3 178 0 5 285 6 0 5 200 1 13 40 3 1 15 3 1 15 Appendix ...
Страница 248: ...240 Precautions when Processing Test Leads ...
Страница 252: ...244 Index ...
Страница 253: ......
Страница 254: ......
Страница 255: ......
Страница 256: ......