79
Overview
Graph display (display graph switching)
Waveform graph display
Displays the waveform graph only.
1
4
5
2
3
1
Number of applied pulses and
peak voltage display
[P:XX/XX]
: Number of pulses that have been applied/Total number of
pulses
[xxxV]
: Max. peak voltage value of the response waveform
2
Waveform judgment area and
LC and RC values operation
range bar
Blue: AREA judgment area
Green: DIFF judgment area
Purple FLUTTER judgment area
Gray: LAPLACIAN judgment area
Yellow (×2): LC and RC value operation areas
3
Response waveform
Yellow: Master waveform
Blue: Test waveform (Failures are displayed in red)
4
Discharge judgment threshold
value bar
Displays the discharge judgment threshold value.
5
Discharge amount graph
Discharge amount graph (Parts where the discharge amount
exceeds the threshold value are displayed in red)
Note: Displayed only if the Discharge Detection Upgrade is enabled.
Reference: “5.4 Discharge Judgments (When Incorporating Model
6
Implementing
Tests
Содержание ST4030
Страница 2: ......
Страница 14: ...6 Usage Notes ...
Страница 32: ...24 Inspecting before Measurement and Verifying Operations ...
Страница 42: ...34 Table Copying ...
Страница 108: ...100 Setting Dielectric Break down Judgment Conditions ...
Страница 132: ...124 Terminal Open Error Setting ...
Страница 198: ...190 Editing Files and Folders ...
Страница 214: ...206 Interface Specifications ...
Страница 237: ...229 Dimensions Diagram 15 4 Dimensions Diagram Unit mm 215 1 348 3 178 0 5 285 6 0 5 200 1 13 40 3 1 15 3 1 15 Appendix ...
Страница 248: ...240 Precautions when Processing Test Leads ...
Страница 252: ...244 Index ...
Страница 253: ......
Страница 254: ......
Страница 255: ......
Страница 256: ......