58
LC and RC Values Area Judgments [LCRC AREA]
Auto creation of the PASS judgment area
The PASS judgment area is created automatically from the imported LC and RC master values.
(Measurement screen)
[MODE]
>
[SETTING]
3
2
1
4
1
Tap
[CREATE]
(on the measurements screen).
2
Tap
[AREA TYPE]
to set the shape of the area when the PASS judgment area is created
automatically.
HI-LO
Creates a rectangular PASS judgment area from the maximum and minimum LC and RC
master values.
Used when the master workpiece’s LC and RC values are distributed in close proximity.
FIT
Creates a banded PASS judgment area that covers the LC and RC master values.
Used if the LC and RC values distribution is banded by the rotor position (angle) with a motor
that has been combined with a rotor.
Rectangular PASS judgment area
[HI-LO]
Banded PASS judgment area
[FIT]
3
Set the margin when creating the PASS judgment area automatically.
[AREA TYPE]
[HI-LO]
: Adds the result of multiplying the LC and RC average values by the margin rate to the
upper and lower limit values.
[FIT]
: Adds the result of multiplying the length of the long and short sides of the banded area
by the margin rate to the outside of the long and short sides, respectively. Since a
value of 100% will cause the original lengths to be added on both sides, the long and
short sides will be 3 times their original length.
Содержание ST4030
Страница 2: ......
Страница 14: ...6 Usage Notes ...
Страница 32: ...24 Inspecting before Measurement and Verifying Operations ...
Страница 42: ...34 Table Copying ...
Страница 108: ...100 Setting Dielectric Break down Judgment Conditions ...
Страница 132: ...124 Terminal Open Error Setting ...
Страница 198: ...190 Editing Files and Folders ...
Страница 214: ...206 Interface Specifications ...
Страница 237: ...229 Dimensions Diagram 15 4 Dimensions Diagram Unit mm 215 1 348 3 178 0 5 285 6 0 5 200 1 13 40 3 1 15 3 1 15 Appendix ...
Страница 248: ...240 Precautions when Processing Test Leads ...
Страница 252: ...244 Index ...
Страница 253: ......
Страница 254: ......
Страница 255: ......
Страница 256: ......