Connecting the Test Lead (on the ST4030)
2.1 Connecting the Test Lead (on the ST4030)
Connect the test lead to the voltage output terminal on the instrument.
CAUTION
• To prevent damage to the coaxial connectors or joint, make sure to unlock first, and
grasp the
coaxial connector and pull it out.
• Model L2250 Clip Type Lead’s maximum rated voltage is 3300 V AC peak. To test
higher voltages, use the L2252 Unprocessed Lead Cable.
• Model L2252 requires customer modification. To prevent electric shock and equipment
damage when modifying the lead, see “15.10 Precautions when Processing Test
Rear
Red
Black
1
Align the groove on the test lead’s coaxial cable to the lock pin on the voltage output
terminal on the ST4030 side and insert.
2
Rotate the coaxial connector on the test lead to the right to lock it.
When removing, rotate the coaxial connector on the connection cable to the left and then pull out.
Содержание ST4030