56
Replacing the Power Fuse
Replaceable Parts and Operating Lifetimes
The characteristics of some of the parts used in the product may deteriorate with extended use.
To ensure the product can be used over the long term, it is recommended to replace these parts on a
periodic basis. When replacing parts, please contact your authorized Hioki distributor or reseller.
The service life of parts varies with the operating environment and frequency of use.
Parts are not guaranteed to operate throughout the recommended replacement cycle.
Part name
Recommended replacement cycle
Electrolytic Capacitors
Approx. 10 years
LCD backlight
Approx. 20,000 hours
Relay
Approx. 1 million operations
6.4 Replacing the Power Fuse
WARNING
•
To avoid electric shock, turn off the power switch and disconnect the connection
cables before replacing the fuse.
•
Replace the fuse only with one of the specified type, characteristics, rated current,
and rated voltage. Do not use fuses other than those specified (especially, do not use
a fuse with higher-rated current) or do not short circuit and use the fuse holder. Doing
so may damage the device and result in bodily injury.
Fuse type: MF60NRF-8A f6.4 mm×30 mm Normal-acting fuse
Prepare: Phillips head screwdriver
Fuse holder
1
Turn off the power switch and disconnect the power cord.
2
Remove the fuse holder using the Phillips head screwdriver.
3
Remove the power fuse from the fuse holder.
Fuse:
φ
6.4 mm 30 mm,
MF60NRF-8A f6.4 mm 30 mm Normal-acting fuse
4
Insert a new power fuse with the specified rating into the fuse holder.
5
Insert the fuse holder into its original location and tighten the screw. (Use the Phillips
head screwdriver)
This completes the process of replacing the power fuse.