6.2 Replacing the Power Fuse
47
6.2
Replacing the Power Fuse
• To avoid electric shock, turn off the power switch and disconnect the
connection cables before replacing the fuse.
• Replace the fuse only with one of the specified characteristics and volt-
age and current ratings. Never use unspecified fuses and never use the
device after the fuse holder has shorted. This will damage the device and
cause injury.
Fuse type: MF60NRF-8A
φ
6.4 mm × 30 mm Normal-acting fuse
Screwdriver
Fuse holder
Fuse:
φ
6.4 mm × 30 mm
Replacing the Power Fuse
Rear panel of the device
Tools to Prepare: Phillips head screwdriver
1
Turn off the power switch and
disconnect the power cord.
2
Loosen the fuse holder screw
with the Phillips head screw-
driver and remove.
3
Replace the power fuse with a
fuse of the designated rating.
4
Insert the fuse holder into its
original location and tighten the
screw.
Removing the Fuse Holder