6.3 Functions
51
6.3 Functions
Measurement value indication
Function
Selects displayed value.
Default state
Resistance
Settings
Resistance * / Current
*The resistance is calculated from the set measurement voltage and current value.
Measurement speed
Function
Selects the measurement speed.
Default state
SLOW2
Setups
FAST / MED (medium) / SLOW / SLOW2
Measurement ranges
Current
1 pA to 1 mA
Resistance
1 × 10
2
Ω
to 1 × 10
15
Ω
Current measurement range switching
Function
Selects the current measurement range.
Default state
AUTO, 10
μ
A
Setups
Functions : HOLD / AUTO
Ranges : 100 pA / 1 nA / 10 nA / 100 nA / 1
μ
A / 10
μ
A / 100
μ
A / 1 mA
Trigger delay time
Function
Fix time between inputting trigger signal and starting measurement.
Default state
0 ms
Setup ranges
0 ms to 9999 ms (1 ms resolution)
Averaging
Function
Carry out averaging procedure of measurement values
Default state
ON, 1
Setups
OFF (No averaging) / ON (Required setting number of times for averaging) / AUTO
(Number of times for averaging is automatically determined)
Number of times (in case “ON” setting): 1 to 255
Averaging method
Moving average
Power source frequency
Function
Sets the power source frequency.
Default state
50 Hz
Setups
50 Hz / 60 Hz