4.4 Configuring the Communications Protocol
23
Setting the address
Communication conditions
Remote switching requests
Send the “
RMT
” command from the RS-232C interface.
4.4 Configuring the Communications Protocol
Configuring GP-IB Interface Communications
1
Press and hold the scroll keys (
PAGE UP
▲
/
PAGE DOWN
▼
) on the front of the instru-
ment for about 7 seconds. (The address can be set from the P1, P2, or P3 screen.)
■
Screen P1: Measured value display
3
When finished making the setting, turn off the instrument.
4
Turn on the instrument.
The instrument will revert to the initial screen, and the GP-IB address will be set to
the selected address.
2
Using the scroll keys (
PAGE UP
▲
/
PAGE DOWN
▼
), set the desired address.
(Valid setting range: 1 to 30)
Configuring RS-232C Interface Communications
Baud rate
38400 bps
Parity
none
Stop bit
1 bit
Data
8 bits
Flow control
none
RS-232C interface communications will not be available until the “
RMT
” com-
mand is sent.