2.3 Connecting the Measurement Cables
13
Connection Methods
2.3 Connecting the Measurement Cables
To avoid damage to the instrument, do not apply voltage to measurement termi-
nals.
Front panel
1
Connect the measurement cables to
each channel’s current input terminal.
2
Connect the voltage output cables to
each channel’s voltage output terminal.
Current input terminals
Voltage output terminals
• The current input terminals incorporate a two-tiered design with both center
and outer conductors. The center conductors are connected to measurement
input, while the outer conductors are connected to guard signals.
• Because the instrument performs high-sensitivity current measurement, noise
occurring on the measurement cables may prevent measured values from sta-
bilizing. Use low-noise shielded measurement cables that meet HIOKI’s speci-
fications.
For more information about measurement cables and voltage output cables,
please contact your dealer or HIOKI representative.