Chapter 13 Specifications
271
13
GP-IB interface (RM3545-01 only)
Communication
contents
Remote control
Device address
0 to 31
Delimiter
LF/ CR+LF
Default setting
Device address: 1, Delimiter: LF
Miscellaneous
Conforms to IEEE 488.2
Interface Functions
SH1 All Source Handshake functions are supported.
AH1 All Acceptor Handshake functions are supported.
T6
Basic talker functions are supported.
Serial poll function are supported.
No talk-only mode.
The talker cancel function with MLA (My Listen Address) is supported.
L4
Basic listener functions are supported.
No listen-only mode.
The listener cancel function with MTA (My Talk Address) is supported.
SR1 All Service Request functions are supported.
RL1
All Remote/Local functions are supported.
PP0
No Parallel Poll function.
DC1 All Device Clear functions are supported.
DT1
All Device Trigger functions are supported.
C0
No Controller functions are supported.
Communications functionality
Remote function
During remote operation by USB, RS-232C or GP-IB, all front panel key oper-
ations are disabled. Remote operation is canceled as follows:
• LOCAL key, Reset, At power-on
• By USB, RS-232C or GP-IB
:SYSTem:LOCal
command
• By GP-IB
GTL
command
Communications
monitor function
Displays the send/receive status of commands and queries.
Setting: ON/ OFF
Data output function
During INT trigger source operation, measured values are output at TRIG sig-
nal or
ENTER
key input.
During EXT trigger source operation, measured values are automatically out-
put each time measurement completes.
(USB keyboard mode is available during INT trigger source use only.)
ON/ OFF
Memory function
Measured values stored in the instrument’s memory are sent at once.
(The memory function is automatically turned off when using the MUX mea-
surement terminal setting.)
No. of memory: 50 (volatile memory, no backup)
ON/ OFF
Default setting
Communications monitor function: OFF, Data output: OFF,
Memory function: OFF
Содержание RM3545
Страница 2: ......
Страница 4: ......
Страница 26: ...Operating Precautions 18...
Страница 42: ...1 5 Checking the Measurement Target 34...
Страница 54: ...2 6 Pre Operation Inspection 46...
Страница 70: ...3 5 Checking Measured Values 62...
Страница 226: ...10 6 Supplied Connector Assembly 218...
Страница 290: ...Chapter 13 Specifications 282...
Страница 312: ...14 4 Disposing of the Instrument 304...
Страница 349: ...Appendix 18 Outline Drawing A37 Appendix Appendix 18 Outline Drawing...
Страница 362: ...Index Index 4...
Страница 363: ......
Страница 364: ......