8.2 Internal Circuitry
147
8
See:
"13.2 Z3003 Multiplexer Unit"(p.277)
(1) Measurement targets (wiring order is user-selected)
(2) Measurable range
(3) Contact specifications
Electrical Specifications
4-wire
2-wire
10 locations (when using two Z3003 units, 20 locations)
21 locations (when using two Z3003 units, 42 locations)
Measurement current
Instrument with Z3003: 1 A DC or less
Externally connected device: 1 A DC or less, 100 mA AC or less
Measurement frequency
Externally connected device: DC, 10 Hz to 1 kHz
Contact type
Mechanical relay
Maximum allowable
voltage
33 V RMS and 45.7 V peak or 70 V DC
Maximum allowable
power
30 W (DC) (Resistance load)
Contact service life
4-wire: 50 million cycles, 2-wire: 5 million cycles (reference value)
Содержание RM3545
Страница 2: ......
Страница 4: ......
Страница 26: ...Operating Precautions 18...
Страница 42: ...1 5 Checking the Measurement Target 34...
Страница 54: ...2 6 Pre Operation Inspection 46...
Страница 70: ...3 5 Checking Measured Values 62...
Страница 226: ...10 6 Supplied Connector Assembly 218...
Страница 290: ...Chapter 13 Specifications 282...
Страница 312: ...14 4 Disposing of the Instrument 304...
Страница 349: ...Appendix 18 Outline Drawing A37 Appendix Appendix 18 Outline Drawing...
Страница 362: ...Index Index 4...
Страница 363: ......
Страница 364: ......