4.3 Setting the Measurement Integration Time Option
44
The integration time can be optionally set for each range by selecting FAST, MED, or SLOW mea-
surement speed. Integration time can be set in ms or PLC* units.
* PLC = Power Line Cycle, where one PLC is the period of the power line waveform. At 50 Hz, one PLC = 1/50
th
of a sec-
ond, and at 60 Hz, one PLC = 1/60
th
of a second.
PLC setting units are useful where measurements may be affected by power line noise (high- or low-resistance measure-
ments)
*1. Low Power = Low-Power Resistance Measurement (p. 28)
*2. Two measurements are made within the above integration times.
4.3
Setting the Measurement Integration Time
Option
Default Settings
Range
LOW POWER: OFF (p. 28) *1
LOW POWER: ON (p. 28) *1
Integration Time [INT]
OVC
Integration Time [INT]
OVC
FAST
MED
SLOW
FAST
MED
SLOW
100m
0.5 ms
5.0 ms
1PLC
ON
*2
1000m
0.3 ms
2.5 ms
1PLC
ON
*2
0.5 ms
5.0 ms
1PLC
ON
*2
10
0.3 ms
2.5 ms
1PLC
ON
*2
0.5 ms
5.0 ms
1PLC
ON
*2
100
0.3 ms
3.0 ms
1PLC
OFF
0.3 ms
2.5 ms
1PLC
ON
*2
1000
0.3 ms
3.0 ms
1PLC
OFF
0.3 ms
2.5 ms
1PLC
ON
*2
10k
0.3 ms
3.0 ms
1PLC
OFF
100k
0.5 ms
3.0 ms
1PLC
OFF
1000k
1.5 ms
5.0 ms
1PLC
OFF
10M
2.5 ms
1PLC
1PLC
OFF
100M
1PLC
2PLC
4PLC
OFF
1
Open the Basic Settings screen.
2
Open the Measurement Settings Screen.
The Basic Settings screen appears.
The Measurement Settings Screen
appears.
[MEAS SETTINGS]
(The settings for the current mea-
surement range are displayed.)
1
Selection
2
Содержание RM3542
Страница 2: ......
Страница 6: ...Contents iv ...
Страница 16: ...Operating Precautions 10 ...
Страница 26: ...1 3 Screen Organization 20 ...
Страница 32: ...2 3 Turning the Power On and Off 26 ...
Страница 46: ...3 8 Confirming Faulty Measurements 40 ...
Страница 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Страница 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Страница 90: ...7 3 Printing 84 ...
Страница 172: ...9 10 Device Compliance Statement 166 ...
Страница 190: ...11 4 Disposing of the Instrument 184 ...
Страница 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Страница 214: ...Index Index 4 ...
Страница 215: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......