![Hioki RM3542-50 Скачать руководство пользователя страница 93](http://html1.mh-extra.com/html/hioki/rm3542-50/rm3542-50_instruction-manual_2137361093.webp)
6.4 Auto-Exporting Measurement Values (at End of Measurement) (Data Output Function)
87
6
At the end of measurement, the measurement value is exported to a computer via RS-232C.
See:
"Chapter 9 Communications (RS-232C/ GP-IB Interface)" (p. 109)
6.4 Auto-Exporting Measurement Values (at End
of Measurement) (Data Output Function)
• Set the interface to [RS232C] beforehand. This function is not applicable to the GP-IB Interface.
See:
"9.4 Configuring the Communications Protocol" (p. 113)
• When internal [TRG: INT] triggering is selected, data is exported only when a TRIG signal is applied.
• Executing a :READ? query command exports duplicate measurement values.
• Even when using other queries, ensure that the query response and auto-exporting of measured values do
not overlap.
• The data format for measurement values can be selected as ASCII (default) or BINARY. Transfer time is
minimized when BINARY is selected.
See:
1
Open the Basic Settings screen.
2
Open the Data Memory Settings screen.
3
Enable or disable the function.
4
Return to the Measurement screen.
The Basic Settings screen appears.
The Data Settings screen appears.
[DATA SETTINGS]
1
Selection
2
Selection
1
2
Disables auto-exporting (default).
Enables auto-exporting.
Returns to the setting screen.
Saves setting and return to previous screen.
Discards setting and return to previ-
ous screen.
The confirmation screen appears.
Содержание RM3542-50
Страница 1: ......
Страница 2: ......
Страница 6: ...Table of Contents iv...
Страница 26: ...1 3 Screen Organization 20...
Страница 32: ...2 3 Turning the Power On and Off 26...
Страница 48: ...3 8 Confirming Faulty Measurements 42...
Страница 72: ...4 14 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 66...
Страница 84: ...5 7 Initializing Reset 78...
Страница 94: ...6 4 Auto Exporting Measurement Values at End of Measurement Data Output Function 88...
Страница 206: ...11 4 Disposing of the Instrument 200...
Страница 216: ...Appendix 5 Dimensional diagram A10 Appendix 5 Dimensional diagram Unit mm...
Страница 230: ...Index Index 4...
Страница 231: ...HIOKI E E CORPORATION...
Страница 232: ......
Страница 233: ......
Страница 234: ......