11.3 Error Displays and Solutions
197
11
The following messages are displayed on the screen when the instrument malfunctions or encounters an
abnormal measurement state.
If repair is necessary, contact your authorized Hioki distributor or reseller.
11.3 Error Displays and Solutions
Display
Description
Remedy
+OvrRng/-OvrRng
Out-of-Range (p. 40)
Select the appropriate range.
C.E. Hi
Hi wiring contact error (p. 40)
Check for cable breakage and worn
out probes.
C.E. Lo
Lo wiring contact error (p. 40)
Check for cable breakage and worn
out probes.
C.E. Volt
Voltage Level Monitor Error
(p. 40)
Check for worn out probes.
ERR:001
LOW limit is higher than UPP
limit.
Upper limit is below lower limit.
Check the comparator settings
(p. 36).
ERR:002
[When the Low-Power Resis-
tance Measurement is OFF]
Exceeding range. (From 0
Ω
to
120 M
Ω
)
[When the Low-Power Resis-
tance Measurement is ON]
Exceeding range. (From 0
Ω
to
1200
Ω
)
Comparator input is out of range.
(out-of-range reference value,
such as 1000 M
Ω
)
Check comparator settings (p. 36).
ERR:003
Setting monitor error. (COMP)
The comparator settings are dif-
ferent from the other instrument.
Make the comparator settings the
same (p. 59).
ERR:004
Setting monitor error. (SPEED)
The SPEED settings are different
from the other instrument.
Make the SPEED settings the same
(p. 59).
ERR:011
Zero adjustment error. Offset
value exceeds 10
Ω
.
Out of zero-adjust range
Check the zero-adjustment proce-
dure (p. 34).
ERR:021
Probe short error
Probe short-circuit
Check if the connections are correct
(p. 57).
ERR:031
Command error
Command Error
Check if the commands (p. 127) are
correct.
ERR:032
Execution error
Execution Error
Check the last command and instru-
ment status (p. 127).
ERR:033
RS-232C communication error
I/F Communications Error
Check communication settings and
wiring.
ERR:034
Setting monitor communica-
tion error
Settings monitor communications
error
Check the setting monitor settings
and wiring (p. 59).
ERR:041
Line frequency detection error
Error detecting the line frequency
Manually set the frequency to match
the power line (p. 71).
ERR:042
Clock error
Reset?(16-01-01 00:00:00)
Press F1 Key.
The clock is not set, so pressing
F1[OK] displays the default time
as 16-01-01 00:00:00.
The back-up battery needs to be re-
placed.
Contact your authorized Hioki dis-
tributor or reseller.
ERR:101
Hardware error
(Main CPU ROM)
Hardware error
Repair is required.
ERR:102
Hardware error
(Main CPU RAM)
Hardware error
Repair is required.
ERR:103
Hardware error (SRAM)
Hardware error
Repair is required.
ERR:104
Hardware error
(Adjustment data)
Hardware error
Repair is required.
Содержание RM3542-50
Страница 1: ......
Страница 2: ......
Страница 6: ...Table of Contents iv...
Страница 26: ...1 3 Screen Organization 20...
Страница 32: ...2 3 Turning the Power On and Off 26...
Страница 48: ...3 8 Confirming Faulty Measurements 42...
Страница 72: ...4 14 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 66...
Страница 84: ...5 7 Initializing Reset 78...
Страница 94: ...6 4 Auto Exporting Measurement Values at End of Measurement Data Output Function 88...
Страница 206: ...11 4 Disposing of the Instrument 200...
Страница 216: ...Appendix 5 Dimensional diagram A10 Appendix 5 Dimensional diagram Unit mm...
Страница 230: ...Index Index 4...
Страница 231: ...HIOKI E E CORPORATION...
Страница 232: ......
Страница 233: ......
Страница 234: ......