9.7 Message Reference
161
9
Set and Query Measurement Current Mode
Syntax
Command
:RESistance:CURRent
<Range>
,
<PULSe/ CONTinuous>
Query
Response
:RESistance:CURRent?
<Range>
<PULSE/ CONTINUOUS>
Command
:LPResistance:CURRent
<Range>
,
<PULSe/ CONTinuous>
Query
Response
:LPResistance:CURRent?
<Range>
<PULSE/ CONTINUOUS>
<Range> = (:RESistance ) RNG100MIL/ RNG1000MIL/ RNG3/ RNG10/ RNG100/
RNG300/ RNG1000/ RNG10K/ RNG30K/ RNG100K/ RNG300K/ RNG1000K/
RNG3MEG/ RNG10MEG/ RNG30MEG/ RNG100MEG
(:LPResistance ) RNG1000MIL/ RNG3/ RNG10/ RNG100/ RNG300/
RNG1000
<PULSe> = The measurement current is applied as a pulse only during the measurement.
<CONTinuous> = The measurement current is applied continuously.
Description
Selects the method of current application. Select (CONTinuous) when measuring objects
that require time for stabilization after applying the measurement current.
Note
When using the Contact Improver function (set to HOLD or PULSE), the measurement
current is applied only as a pulse, even if CONTinuous has been selected. Therefore, to
measure with a continuous current, the Contact Improver function must be disabled (set to
OFF) (p. 160).
Set and Query contact check operation
Syntax
Command
:RESistance:CONTactcheck
<Range>
,
<1/ 0/ ON/ OFF>
Query
Response
:RESistance:CONTactcheck?
<Range>
<ON/ OFF>
Command
:LPResistance:CONTactcheck
<Range>
,
<1/ 0/ ON/ OFF>
Query
Response
:LPResistance:CONTactcheck?
<Range>
<ON/ OFF>
<Range> = (:RESistance ) RNG100MIL/ RNG1000MIL/ RNG3/ RNG10/ RNG100/
RNG300/ RNG1000/ RNG10K/ RNG30K/ RNG100K/ RNG300K/ RNG1000K/
RNG3MEG/ RNG10MEG/ RNG30MEG/ RNG100MEG
(:LPResistance ) RNG1000MIL/ RNG3/ RNG10/ RNG100/ RNG300/
RNG1000
Description
Enables/disables the Contact Check function.
Содержание RM3542-50
Страница 1: ......
Страница 2: ......
Страница 6: ...Table of Contents iv...
Страница 26: ...1 3 Screen Organization 20...
Страница 32: ...2 3 Turning the Power On and Off 26...
Страница 48: ...3 8 Confirming Faulty Measurements 42...
Страница 72: ...4 14 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 66...
Страница 84: ...5 7 Initializing Reset 78...
Страница 94: ...6 4 Auto Exporting Measurement Values at End of Measurement Data Output Function 88...
Страница 206: ...11 4 Disposing of the Instrument 200...
Страница 216: ...Appendix 5 Dimensional diagram A10 Appendix 5 Dimensional diagram Unit mm...
Страница 230: ...Index Index 4...
Страница 231: ...HIOKI E E CORPORATION...
Страница 232: ......
Страница 233: ......
Страница 234: ......