9.7 Message Reference
159
9
(11)Measurement Conditions
Set and Query Delay 2 (Delay time after the measurement current has been
applied)
Syntax
Command
:RESistance:DELay2
<Range>
,
< Delay 2 time>
Query
Response
:RESistance:DELay2?
<Range>
<Delay 2 time>
Command
:LPResistance:DELay2
<Range>
,
< Delay 2 time>
Query
Response
:LPResistance:DELay2?
<Range>
<Delay 2 time>
<Range> = (:RESistance ) RNG100MIL/ RNG1000MIL/ RNG3/ RNG10/ RNG100/
RNG300/ RNG1000/ RNG10K/ RNG30K/ RNG100K/ RNG300K/ RNG1000K/
RNG3MEG/ RNG10MEG/ RNG30MEG/ RNG100MEG
(:LPResistance ) RNG1000MIL/ RNG3/ RNG10/ RNG100/ RNG300/
RNG1000
<Delay2 time> = 0.0000 to 0.10000 (NR2) [sec]
Example
:RES:DEL2 RNG1000MIL,0.003
:RES:DEL2? RNG10
0.0050
Set and Query Integration Time (multiples of power line cycles)
Syntax
Command
:RESistance:NPLCycles
<Range>
,
<Speed>
,
< Integration
time (NPLC)>
Query
Response
:RESistance:NPLCycles?
<Range>
,
< Speed>
<Integration time (NPLC)>
Command
:LPRESistance:NPLCycles
<Range>
,
<Speed>
,
< Integration time (NPLC)>
Query
Response
:LPResistance:NPLCycles?
<Range>
,
< Speed>
<Integration time (NPLC)>
<Range> = (:RESistance ) RNG100MIL/ RNG1000MIL/ RNG3/ RNG10/ RNG100/
RNG300/ RNG1000/ RNG10K/ RNG30K/ RNG100K/ RNG300K/ RNG1000K/
RNG3MEG/ RNG10MEG/ RNG30MEG/ RNG100MEG
(:LPResistance ) RNG1000MIL/ RNG3/ RNG10/ RNG100/ RNG300/
RNG1000
<Speed> = SLOW/ MEDium/ FAST
<Integration time (NPLC)> = 0.0100 to 6.0000 (NR2)
Description
Sets the integration time as an integer multiple of the power line cycles (PLC).
Note
•
For high-resistance measurements, hum noise is suppressed and stable measurement
values are acquired by setting an integer multiple of power line cycles.
•
The values are converted to seconds, and values over 100 ms are set to 100 ms.
•
When the integration time is acquired with a query, the instrument internal variable is
converted and output.
In some cases, the returned value may not match the set value.
Содержание RM3542-50
Страница 1: ......
Страница 2: ......
Страница 6: ...Table of Contents iv...
Страница 26: ...1 3 Screen Organization 20...
Страница 32: ...2 3 Turning the Power On and Off 26...
Страница 48: ...3 8 Confirming Faulty Measurements 42...
Страница 72: ...4 14 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 66...
Страница 84: ...5 7 Initializing Reset 78...
Страница 94: ...6 4 Auto Exporting Measurement Values at End of Measurement Data Output Function 88...
Страница 206: ...11 4 Disposing of the Instrument 200...
Страница 216: ...Appendix 5 Dimensional diagram A10 Appendix 5 Dimensional diagram Unit mm...
Страница 230: ...Index Index 4...
Страница 231: ...HIOKI E E CORPORATION...
Страница 232: ......
Страница 233: ......
Страница 234: ......